USER MANUAL DO-LBL-06035(L)
| MyoVista® Wavelet ECG (
wav
ECG™) 12-Lead Cardiac Testing Device
Service – Maintenance & Cleaning
Periodic Inspection
Visually inspect the MyoVista Device every 6 months or as required by facility protocol .
Warning: If service is required, it should be performed only by authorized service personnel.
Seasonal Safety Check
Clean the power adapter plug at least once a year . An excessive amount of dust on the plug might
cause a fire .
Note: For proper performance and safety, have the device inspected and tested annually by
authorized service personnel. Inspection and testing should be performed only by authorized service
personnel trained in maintenance of the MyoVista Device.
Annual Inspection
Perform the following safety checks at least every 12 months and record the data per your facilities
protocol . If the device is not working properly or fails any of the following conditions, it must be
repaired:
• Inspect the equipment and accessories for mechanical and functional damage .
• Inspect the relevant safety labels for legibility .
• Verify that the device functions properly, as described in this User Manual .