USER MANUAL DO-LBL-06035(L)
| MyoVista® Wavelet ECG (
wav
ECG™) 12-Lead Cardiac Testing Device
Safety Summary
Training
This User Manual provides information about patient setup, performing a test, interpreting the
results, and using the software . It is intended as a supplement to, not a substitute for, thorough
product training .
Medical Device Classification
The device is classified as follows per ANSI/AAMI/EN/IEC 60601-1:2005/(R)2012 .
Category
Classification
Protection against electric shock
Class I externally powered ME equipment . Internally
powered ME equipment .
Degree of protection against electrical shock
Defibrillation-Proof Type CF Applied Part .
Protection against harmful ingress of water or
particulate matter
The device is classified IP00 (non-protected) for
harmful ingress of particulate matter and water .
Methods of sterilization
N/A
Suitability for use in an oxygen-rich environment
The device is not suitable for use in the presence of
flammable vapors or gases .
Mode of operation
Continuous operation