3 . To back up the database before restoring factory defaults, select the Backup icon . Otherwise,
select the Skip icon to go directly to the factory reset operation . Selecting the Cancel icon closes
the screen without restoring the factory default settings .
– If you select the Backup icon, you are prompted to complete a database backup . After this
function is performed, you are prompted to proceed and erase data or cancel the operation .
See “Backing Up a Database” on page 76 .
– If you select the Skip icon, you are prompted to proceed and either erase data or cancel the
operation .
• If you select the Cancel icon, the factory reset operation ends, the following message
appears, and you should select OK to return to the Update/Factory Reset screen:
The Factory Reset operation has been aborted.
• If you select the Proceed and Erase Data icon, the device reverts to factory default settings .
• A pop-up window with a warning will be displayed warning the user not to remove power
from the device until the update/factory reset process completes . Click on OK .
4 . The system will start powering down and will go through the reset sequence . The device will boot
the MyoVista home screen after reset . The process takes about 2 minutes .
Note: If needed, contact customer service for instructions on how to reset the device
Restoring/Updating Device Reference Materials
(This is accessible only by Administrator roles .)
1 . Select the Settings Menu, then click on Update/Factory Reset icon and then click on Update
Manuals icon . A pop up will appear stating “The Reference Materials on the Home tab will be
updated .” Click on Continue .
2 . From the USB drive, select the User Manuals update file (MyoVista_DG-200_Manuals_Update_
Package_Rev-* .upm) and click OK . Let the MyoVista Update Manuals process complete .
3 . Once completed, a pop up will indicate successful update of the user manuals . Click on OK and
then click on Cancel in the Update/Factory Reset tab .
* The letter that appears at this location in the filename indicates the revision of the document