108
USER MANUAL DO-LBL-06035(L)
| MyoVista® Wavelet ECG (
wav
ECG™) 12-Lead Cardiac Testing Device
Appendix
A . Technical Data
Device Specifications
Specification
Description
Dimensions (H X W X D)
11 .75 in . x 16 .75 in . x 3 .0 in . excluding stand or accessories
Weight
8 .4 lbs . including battery
Patient cable connection
DB-15 ECG with proprietary pin-outs
Ports
USB (4), DC Power Input, HDMI, LAN
Materials
Bayblend FR3010 - Polymer blend based on bisphenol A -polycarbonate/acrylonitrile-
butadiene-styrene copolymer (PC/ABS), flame-retarded (low halogen content, free from
chlorine and bromine)
Disposal
According to WEEE: Directive 2012/19/EU - Device, accessories, and packaging waste must
be disposed of properly at end of each usage . Follow Local Ordinances and Regulations
for disposal .
Built-In Monitor
Size
Effective display area 15 .6 in .
Resolution
1920 x 1080 dots per inch (DPI)
Controls
Touchscreen and membrane key panel
Memory
Storage
Patient Test Records: > 50,000
Electrical Specifications
Specification
Description
Power Supply
Voltage
100~240 VAC, 50/60 Hz
Power consumption
48 Watts max
Battery
Operation with built-in rechargeable battery
Battery (new battery with full charge)
Capacity
Approximately 1 hour normal use . Approximately 2 hours standby
Battery life
Normal operating conditions: 1 yr .
Recharging time
90% Capacity: approximately 4 .5 hrs . / 100% Capacity: approximately 10 hrs .
Line frequency filter
Distortion-free suppression of superimposed 50 or 60 Hz sinusoidal interferences by
adaptive digital filtering
Battery (new device, unpackaged, in storage)
Battery life
Normal storage conditions: 6 months
Environmental Specifications - MyoVista Wavelet ECG
Specification
Description
Temperature
Operating
+5°C to +40°C (41°F to 104°F)
Storage
-20°C to +50°C (-4°F to 122°F)
Battery (new battery with full charge)
Operating
≤80% RH non-condensing
Storage
≤80% RH non-condensing