WE2107
A2087-1.3 en/de
HBM
4
Displays
and control functions
Measured
value
lies in
range of
+1/4 d
Battery display
(when supply voltage
too low)
Display of set scale type
and parameter settings
On display of net value, simultaneously with
Net
Units of
measure
ment
Hold
function
activated
Saved manual tare value is used
When the net
value is
displayed
ENTER
UP
MENU
CE
NEXT
WE2107
Lb
kg
tg
<
>
0
PT
Net
Tare
Hold
2
1
G/N
T
F1
F2
Adhesive label with hidden switch
(key) for the calibration menu
Insertion strip for calibration data, device name, etc.
Display window with a 5‐figure
digital display and special symbols
Control keys for scale and
menu functions
The calibration menu key can be accessed with a thin object. After calibration,
the opening must be sealed with the enclosed adhesive label or, for
legal‐for‐trade applications, with the calibration label. Device calibration is pro
tected in operation and can only be changed when this pushbutton is actu
ated.
Each of the four control keys has a basic function for scale operation, which is
identified by a symbol on the key.
“G/N”
Key for toggling between the gross and net display.
“>T<”
Key for taring and switching to the net display.
“F1”
Function key, defined by user in parameter menu.
“F2”
Function key, defined by user in parameter menu.
The labeling above the keys indicates the second function of the keys during
parameter input (menu guidance).