37
WE2107
A2087-1.3 en/de
HBM
Main menu
2nd level 3rd level Sub‐menu explanation
Prt_S
timE
Print protocol settings: xx:xx (real‐time clock Hours : Minutes)
modE
Time mode selection: 24 h / 12 h am / 12 h pm
dAtE
dAY
Day: xx
month
Month: xx
YEAr
Year: xx
Init
ESC11
ESC sequence 1, byte 1: 0...255 (0=OFF)
ESC12
ESC sequence 1, byte 2: 0...255
ESC13
ESC sequence 1, byte 3: 0...255
ESC14
ESC sequence 1, byte 4: 0...255
ESC15
ESC sequence 1, byte 5: 0...255
ESC21
ESC sequence 2, byte 1: 0...255 (0=OFF)
ESC22
ESC sequence 2, byte 2: 0...255
ESC23
ESC sequence 2, byte 3: 0...255
ESC24
ESC sequence 2, byte 4: 0...255
ESC25
ESC sequence 2, byte 5: 0...255
idEnt
nb
Print protocol counter: 0...99999 (0=OFF)
Frt
E_Ch
Number of spaces in each line: 0...99
E_Ln1
Number of empty lines before printing: 0...99
E_Ln2
Number of empty lines after the protocol: 0...99
inPut
InP_1
oFF / tArE / Prt / tilt (OM, Page 54)
InP_2
oFF / tArE / Prt / Loc_P
t_dLY
Tilt delay time: (0...99) x 100 ms
buttn
F1
_F1_ (short), function key assignment, see OM, Page 56
F1_L
_F1_ (long) ”
F2
_F2_ (short) ”
F2_L
_F2_ (long) ”
tESt
diSPL
Test display, LCD segments ON/OFF (via “NEXT” button)
UArt
Transmission of 55Hex... (see OM, Page 58)
d_IO
Test digital inputs/outputs (see OM, Page 58)
EEPro
Checksum check in EEPROM: 0 = error, 1 = OK
buttn
Button test (see OM, Page 58)
ScALE
Funct
Scale function: StAnd / FILL1 / FILL2 / FILL3
ACCES
Menu access levels: 0...4
LEGAL
Legal‐for‐trade operation: OFF / OIML / NTEP
AdJ
SEtUP
UnIt
Physical unit: OFF / g / kg / t / lbs
AZEro
Zero on startup: OFF / 2 / 5 / 10 / 20 / 50 %
ZtrAc
Automatic zero tracking: OFF / ON (0.5d/s)
StiLL
Standstill monitoring: OFF / +- 0.5 / 1 / 2 / 5 d/s
rES
Increment: 1 / 2 / 5 / 10 / 20 / 50 d
Point
Decimal point: 0=xxxxx / 1=xxxx.x / 2=xxx.xx / 3=xx.xxx / 4=x.xxxx
CAP
Max. capacity (weighing range): 100...99999
rAnG1
Changeover point dual‐range scale (0 = single‐range balance)
rAnG2”
Changeover point three‐range scale (0 = single/dual‐range balance)
CAL
Calibration weight: 100...99999
EA_CL
Gravitational acceleration at place of calibration: 9.7000...9,9000
EA_UL
Gravitational acceleration at installation location: 9.7000...9,9000
inPut
ZEro
Value when scale is empty (but with initial load): 0.0000...2.0000
mV/V
SPAn
Value at maximum capacity (but with initial load): 0.0000...2.0000 mV/V
MEAS
ZEro
Actual display when scale is empty (zero point adjustment with
“ENTER”)
SPAn
Actual display with calibration weight (span adjustment with “ENTER”)
Lin
diSP1
Display value 1
VAL1
Measured value 1 for dISP1
diSP2
Display value 2
VAL2
Measured value 2 for dISP2
F_AdJ
dEFLt
Restore the factory settings: YES / NO
oFF
off
Switch off the device with “ENTER”