HBM
5.4.2 Basic scale function, menu access selection and legal‐for‐trade
operation
This function can only be executed when the hidden pushbutton (for calling
the parameter menu) is pressed.
Access
level
Main
menu
level
1
Second
menu
level
2
Third
menu
level
1 2
Type
Sub‐menu explanation
1 2
(flashing) = parameter input /
selection
0
“ScALE
³
Basic scale functions
Funct”
S
Basic scale function selection:
StAnd – non‐automatic
scale/
FILL1 - Dosing function/
FILL2 - Tank weighing
/Fill 1 /
FILL3 - Tank weighing
/Fill 2 /
ACCES”
S
Menu access level selection: 0
…
4
LEGAL”
S
Legal‐for‐trade operation:
OFF / OIML / NTEP
See also OM, Chapter 12
Parameter access types: D – display only, S – individual item selection, I – parameter input; M – measurement
Funct parameter:
Standard (StAnd): non‐automatic scale (NAWI) with/without limit value
switches, counting scale
FILL1,2,3:
Filling and dosing function
(see Chapter 20, Operating Manual, Part 1)
ACCES parameter:
The various menu levels have so‐called access levels. Access to change
parameters is defined with the “ScALE
³
ACCES” parameter. The lowest
level is zero. This feature is used to define user access to the parameters.
The “ScALE
³
ACCES” parameter is protected by the hidden pushbutton.
The first parameter description column shows the relevant access level of
each item in the main menu.
ACCES parameter
Access levels
0
0 only
1
0 and 1
2
0 to 2
3
0 to 3
4
All (standard factory setting)