WE2107
28
A2087-1.3 en/de
HBM
- Input the zero value (mw0) at ZEro
- Input the nominal value (mw1) at SPAn
Scales with several load cells supply the nominal output signal (2 mV/V), if the
weight that applied is the sum of all the maximum load cell capacities.
4 load cells connected in parallel, each at 20 t, sensitivity 2 mV/V.
Maximum capacity of the scale is 60 t.
³
Maximum load cell capacity: = 4
∗
20t = 80 t
³
Weighing range: = 2 mV/V
∗
60t / 80 t = 1.5 mV/V
Adjustment in mV/V does not achieve the same accuracy as adjustment with
calibration weights and so is only used for non legal‐for‐trade applications,
where required accuracy is less.
6.5
Multi‐range scale
The WE2107 can be operated as a single‐range or multi‐range scale.
Two parameters are available for this in the “AdJ
³
SEtuP” parameter menu.
The weighing range (CAP) must be set before this setting.
Scale type
Parameters
Single‐range balance
rAnG1 = rAnG2 = 0
Dual‐range balance
0 < rAnG1 < CAP and rAnG2 = 0
Three‐range scale
0 < rAnG1 < rAnG2 < CAP
The increment for two/three weighing ranges is automatically derived from the
rES parameter, where this parameter always describes the increment of the
first measuring range:
“rES” = 1d
³
Increment range 2 = 2d
³
Increment range 3 = 5d
“rES” = 2d
³
Increment range 2 = 5d
³
Increment range 3 = 10d