21
WE2107
A2087-1.3 en/de
HBM
Access
level
Sub‐menu explanation
1 2
(
flashing) = parameter input /
selection
Type
Third
menu
level
1 2
Second
menu
level
2
Main
menu
level
1
rAnG2”
I
Changeover point
for three‐range scale: 0
…
99999 of
nominal (rated) value,
0 = one/two‐range balance
SEtUP
³
CAL”
I
Calibration weight: 100
…
99999,
CAL = CAP = 100 %
EA_CL”
I
Gravitational acceleration
at place of calibration:
9.7000
…
9.9000
EA_UL”
I
Gravitational acceleration
at installation location:
9.7000
…
9.9000
InPut
³
Input known scale
characteristic curve
ZEro”
I
Value when scale is empty, but
with initial load:
±
99999
(0 = standard value = 0 mV/V)
SPAn”
I
Value at calibration weight:
±
99999
(2.0000 = standard value
= 2 mV/V)
MEAS
³
Measure scale characteristic curve
ZEro”
M
Actual display when scale is empty
(OK with “
ENTER”
key)
SPAn”
M
Actual display when calibration
weight on scale
(OK with
“ENTER”
key)
Lin
³
Linearization correction between
ZEro and SPAn
1)
diSP1”
I
Display value 1, condition:
0 < diSP1 < diSP2
VAL1”
I
Measured value 1
for diSP1,
(0 < VAL1 < VAL2)
diSP2”
I
Display value 2, condition:
diSP1 < diSP2 < nom. resolution
(CAP)
VAL2”
I
Measured value 2 for dISP2,
(0 < VAL1 < VAL2 < CAP)
Parameter access types: D – display only, S – individual item selection, I – parameter input; M – measurement
1)
Linearization is switched OFF, when disp1=disp2=val1=val2=0, see Section 11 f Operating Manual (Part 1).