31
WE2107
A2087-1.3 en/de
HBM
Error code
Recovery
Description
42
Dosing time exceeded
Check flow of material when
dosing
43
Dosing start weight too high
Empty scale beforehand
44
Tank filling too low (not enough material for
next batch)
Top up with material
45
Cumulative weight overflow
Use print function to clear
sum (Prt4 or 6, see Chapter
15.5, OM, Part 1)
46
Incorrect multi‐range parameter
Check parameters,
Chapter 10.5 (OM, Part 1)
Communication errors:
Error code
Description
Recovery
16
Several errors simultaneously
17
Input parameter outside input range
See
Description Part 2
18
Unknown command
19
Incorrect password (
DPW
command)
See
Description Part 2
20
Write‐protected parameter
Mandatory calibration
switched on
21
Print time timeout
No standstill within 5s
Print command cleared
22
LFT counter overflow (
TCR
command)
³
Repair
23
COM1 parity / framing error
Check COM interface setting