4.3 Optional templates
In addition to the essential templates required for describing the transducer,
the Standard IEEE 1451.4 also provides for optional templates which allow
the opportunity to store further information about the transducer in the TEDS
memory. Further optional templates can be provided manufacturer-specific by
HBM.
HBM template: Signal Conditioning
When the transducer manufacturer (or user) creates the template, a definition
is entered about the possible settings for which entries may be stored in the
TEDS memory. As an example it is possible to create the template in such a
way that the only value which may be stored in TEDS is the cut-off frequency
for the lowpass filter.
This has the advantage for users who prefer to carry out a manual zero
balance that the Zero Compensation Value parameter will not be overwritten
next time the TEDS memory is imported.
Parameters
Value
(example)
unit
Change
re-
quires
rights
to level:
Explanation
Lowpass Filter Character-
istics
Bessel
USR
Possible entries: Bessel or Butter-
worth
Lowpass Filter Frequency
1.000
Hz
USR
If the TEDS memory contains values
which the measuring amplifier can-
not adopt - because filter settings
are only possible in certain fre-
quency steps - the next closest step
is set up.
High-pass Filter
Frequency
100.000
Hz
USR
If the TEDS memory contains values
which the measuring amplifier can-
not adopt - because filter settings
are only possible in certain fre-
quency steps - the next closest step
is set up.
Zero Compensation Value 480.000m
Phys.
unit
USR
Zero offset
Tare Value
0.000
Phys.
unit
USR
Tare