18
TEDS
A1813-1.1 en
HBM
HBM template:
Freqsensor:
Physical and electrical units are specified before a template is created by the
manufacturer. In addition, a selection must be made during creation between
the variants ”Frequency” or ”Counter”.
It must also be specified which resolution should be used: 24 Bit, 16 Bit or 8
Bit.
Parameters
Value
Unit
Chang
e re-
quires
rights
to
level:
Explanation
Transducer electrical Signal
Type
Pulse Sensor
ID
Minimum Velocity
0.000E+000
rpm
CAL
Physical Measurement pa-
rameters and units are de-
Maximum Velocity
1.000E+004
rpm
CAL
rameters and units are de
fined when the template is
created
Pulse Measurement Type
Frequency
CAL
Measuring mode frequency
measurement or pulse
counting
Minimum Electrical Value
0.0
Hz
CAL
The difference between
these values is the sensi-
tivity according to the HBM
Maximum Electrical Value
60000.0
Hz
CAL
tivity according to the HBM
data sheet, the HBM test
certificate or from calibra-
tion.
Mapping Method
Linear
Cannot be changed
Discrete Signal Type
Bipolar
Input mode:
Contact to ground,
Contact to Power,
Discrete Signal Amplitude
4
Volts
Active Low,
Active High,
Bipolar
Discrete Signal Configuration
single
Evaluation F2 signal;
Single
Double 90 degree
Double phase plus zero
index
Transducer Response Time
1.0000000u
seconds
Of no significance for HBM