16
TEDS
A1813-1.1 en
HBM
HBM template: Inductive Displacement Transducer
The physical unit is defined when the transducer manufacturer (or user)
creates the template. At the time of creating the template it is also necessary
to choose between the options ”Full Precision”, ”mV/V” and ”
µ
V/V” for the
accuracy of the characteristic transducer curve mapped in TEDS. HBM
always opts for ”Full Precision” here, in order to be able to use full digital
resolution. This choice is also recommended to operators who program the
TEDS memory themselves.
Parameters
Value
1)
Unit
Chang
e re-
quires
rights
to
level:
Explanation
Transducer Electrical
Signal Type
Bridge
sensor
ID
Minimum Distance
0.000E+000
mm
CAL
The difference between these val-
ues is the nominal displacement ac
Maximum Distance
5.000E+001
mm
CAL
ues is the nominal displacement ac-
cording to the HBM data sheet.
Minimum Electrical
Value
0.000
V/V
CAL
The difference between these val-
ues is the sensitivity according to
Maximum Electrical
Value
80.000m
V/V
CAL
ues is the sensitivity according to
the HBM data sheet, the HBM test
certificate or from calibration.
Mapping Method
Linear
This entry cannot be changed.
Bridge type
Full
The bridge type.
The following values are available
for selection: ”Quarter” for quarter
bridge, ”Half” for half bridge, ”Full”
for full bridge. Some HBM transduc-
ers can be connected as half
bridges or full bridges according to
choice.
Transducer Response
Time
1.0000000u
sec-
onds
ID
Of no significance for HBM trans-
ducers.
Excitation Level (Nomi-
nal)
2.5
Volts
ID
Nominal excitation voltage accord-
ing to the HBM data sheet.
Excitation Level (Maxi-
mum)
10.0
Volts
ID
Upper limit for the operating range
of the excitation voltage according
to the HBM data sheet.
Excitation Voltage Type
AC (rms)
ID
This entry cannot be changed.
Excitation Frequency
(Nominal)
4801
Hz
ID
Nominal value for the carrier fre-
quency of the excitation voltage ac-
cording to the HBM data sheet.
Excitation Frequency
(Minimum)
4421
Hz
ID
Lower limit of the operating range
for the carrier frequency of the ex-
citation voltage according to the
HBM data sheet.
Excitation Frequency
(Maximum)
5181
Hz
ID
Upper limit of the operating range
for the carrier frequency of the ex-
citation voltage according to the
HBM data sheet.
1)
Typical values for a type WA displacement transducer, measuring range 50 mm, from HBM