HBM template: Inductive Displacement Transducer

The physical unit is defined when the transducer manufacturer (or user)

creates the template. At the time of creating the template it is also necessary

to choose between the options ”Full Precision”, ”mV/V” and ”

µ

V/V”  for the

accuracy of the characteristic transducer curve mapped in TEDS. HBM

always opts for ”Full Precision” here, in order to be able to use full digital

resolution. This choice is also recommended to operators who program the

TEDS memory themselves.

Parameters

Value 

1)

Unit

Chang

e re-

quires

rights

to

level:

Explanation

Transducer Electrical

Signal Type

Bridge

sensor

ID

Minimum Distance

0.000E+000

mm

CAL

The difference between these val-

ues is the nominal displacement ac

Maximum Distance

5.000E+001

mm

CAL

ues is the nominal displacement ac-

cording to the HBM data sheet.

Minimum Electrical

Value

0.000

V/V

CAL

The difference between these val-

ues is the sensitivity according to

Maximum Electrical

Value

80.000m

V/V

CAL

ues is the sensitivity according to

the HBM data sheet, the HBM test

certificate or from calibration.

Mapping Method

Linear

This entry cannot be changed.

Bridge type

Full

The bridge type.

The following values are available

for selection: ”Quarter” for quarter

bridge, ”Half” for half bridge, ”Full”

for full bridge. Some HBM transduc-

ers can be connected as half

bridges or full bridges according to

choice.

Transducer Response

Time

1.0000000u

sec-

onds

ID

Of no significance for HBM trans-

ducers.

Excitation Level (Nomi-

nal)

2.5

Volts

ID

Nominal excitation voltage accord-

ing to the HBM data sheet.

Excitation Level (Maxi-

mum)

10.0

Volts

ID

Upper limit for the operating range

of the excitation voltage according

to the HBM data sheet.

Excitation Voltage Type

AC (rms)

ID

This entry cannot be changed.

Excitation Frequency

(Nominal)

4801

Hz

ID

Nominal value for the carrier fre-

quency of the excitation voltage ac-

cording to the HBM data sheet.

Excitation Frequency

(Minimum)

4421

Hz

ID

Lower limit of the operating range

for the carrier frequency of the ex-

citation voltage according to the

HBM data sheet.

Excitation Frequency

(Maximum)

5181

Hz

ID

Upper limit of the operating range

for the carrier frequency of the ex-

citation voltage according to the

HBM data sheet.

1)

Typical values for a type WA displacement transducer, measuring range 50 mm, from HBM

Содержание TEDS

Страница 1: ...A1813 1 1 en TEDS memory module in transducer Content and editing the memory module Operating Manual...

Страница 2: ......

Страница 3: ...main templates 12 4 1 Basic TEDS 13 4 2 Templates for definition of transducer type 13 4 3 Optional templates 20 4 4 Other templates 23 5 Examples for operating with TEDS 24 5 1 Example 1 24 5 2 Exam...

Страница 4: ...n instructed to carry out the installation commissioning maintenance or repair of the transducer TEDS module and amplifier must have read and understood the Operating Manual and in particular the tech...

Страница 5: ...ir or soldering work on motherboards is prohibited Qualified personnel This module is only to be installed and used by qualified personnel strictly in accordance with the specifications and within the...

Страница 6: ...into its own settings and measurement can then start A special circuit patent applied for makes it possible to use the existing cable leads for transferring the TEDS data This means that the same cabl...

Страница 7: ...contains characteristic transducer data that can be defined by the manufacturer by calibration or by the user They are organized into so called templates which store defined groups of data in a summa...

Страница 8: ...one entry to the next within a template Standard rights USR level This level concerns rights that the operator of the transducer needs in order to change entries which depend on the conditions of use...

Страница 9: ...manufacturer specific it is recommended for compatibility reasons always to arrange the manufacturer specific templates in the TEDS memory after all templates which meet the IEEE 1451 4 standard Other...

Страница 10: ...odule NOTE If the user rights in levels CAL or IDENT are used there is a danger of unintentional overwriting of the central transducer data 3 3 1 Editing TEDS in the HBM view The HBM view is orientate...

Страница 11: ...or low pass filtering tare or zero value This data is optional if no values are entered in TEDS the respective settings of the amplifier are retained when reading in the TEDS data Units conversion HBM...

Страница 12: ...etely new TEDS contents the individual templates must first be generated via the command sequence Edit New New template The minimum requirement is at least one template created that describes the resp...

Страница 13: ...be shown in plain text Version letter ID Possible values a letter from A to Z Version number 7 ID Possible values a number from 0 to 63 The serial number positions that precede the last 7 places for...

Страница 14: ...e 1 999m V V CAL ues is the sensitivity according to the HBM test certificate and from calibration Mapping Method Linear This entry cannot be changed Bridge type Full ID The bridge type The following...

Страница 15: ...Feb Mar Apr May Jun Jul Aug Sep Oct Nov Dec Calibration Initials HBM CAL Initials of the calibrator or calibration laboratory concerned Calibration Period Days 730 days CAL Time before recalibration c...

Страница 16: ...test certificate or from calibration Mapping Method Linear This entry cannot be changed Bridge type Full The bridge type The following values are available for selection Quarter for quarter bridge Ha...

Страница 17: ...e used Format day month year Abbreviations for the months Jan Feb Mar Apr May Jun Jul Aug Sep Oct Nov Dec Calibration Initials HBM CAL Initials of the calibrator or calibration laboratory concerned Ca...

Страница 18: ...eters and units are de fined when the template is created Pulse Measurement Type Frequency CAL Measuring mode frequency measurement or pulse counting Minimum Electrical Value 0 0 Hz CAL The difference...

Страница 19: ...ID Upper limit for the operat ing range of the excitation voltage according to the HBM data sheet Excitation Voltage Type Bipolar DC ID This entry cannot be changed Excitation Current Draw 300 000u A...

Страница 20: ...the lowpass filter This has the advantage for users who prefer to carry out a manual zero balance that the Zero Compensation Value parameter will not be overwritten next time the TEDS memory is import...

Страница 21: ...urve segments 1 Elements CAL 1 Starting domain value of segment of full span 0 CAL Calibration curve polynomial 3 Elements CAL 1 Power of domain value 1 1 0 CAL y a1x1 a2x2 a3x3 Power for the first po...

Страница 22: ...n table 3 Elements CAL contains 3 points of the characteristic curve Elements 1 Domain Calibration Point of full span 0 00 CAL of full span 1 Range Calibration deviation of full span 0 0030 CAL Percen...

Страница 23: ...escribe other transducer types TEDS enabled HBM amplifiers currently recognize the following templates from this group High Level Voltage Output Sensors Transducers with voltage output Resistance Sens...

Страница 24: ...5 AP455i MGC connected to PC via e g USB or Ethernet Operated via PC software catman Easy When starting catmanEasy check that the option Sensor Scan is activated During booting up catman Easy automati...

Страница 25: ...asy component on the Signal processing tab This creates an additional HBM specific template which can then only be interpreted by HBM measuring amplifiers It is ignored by other measuring amplifiers N...

Страница 26: ...o the TEDS memory of the transducer Changes are possible at a later date MGC connected to PC e g USB or Ethernet interface PC has the MGCplus Assistant software included in the MGC scope of delivery a...

Страница 27: ...annel that the TEDS transducer is connected to here Channel 3 In the subsequent context menu select the menu item Edit TEDS This starts the TEDS Editor Step II in the TEDS Editor Click on HBM view Lef...

Страница 28: ...and enter the calibration initials to identify the person or body that implemented the calibration Enter the recalibration deadline following the quality management guidelines at the operator in days...

Страница 29: ...The general unit bar is not supported because of the Standard IEEE 1451 4 However to display the unit bar the additional HBM specific Template Unit Conversion is used This is available by default and...

Страница 30: ...in the column Zero value Tare value or zero using the displayed button NOTE A zero adjustment is generally not stored in the TEDS memory as this is dependent on the installation situation of the tran...

Страница 31: ...2 1 Setting up via the AB22A display control panel Connect the transducer to the measuring device Display the channel to which the TEDS transducer is connected on the display select with the channel...

Страница 32: ...suring device Start the Assistant can also be implemented before connecting the transducer Select the channel to which the TEDS transducer is connected in the Assistant view Transducer see Page 27 Rig...

Страница 33: ...inal value of supply voltage for TEDS data memory V 5 Working range of supply voltage for TEDS data memory V 3 4 6 0 Data transfer protocol As specified in IEEE standard 1451 4 one wire protocol Data...

Страница 34: ...34 TEDS A1813 1 1 en HBM...

Страница 35: ...35 TEDS A1813 1 1 en HBM...

Страница 36: ...45 D 64293 Darmstadt Germany Tel 49 6151 8030 Fax 49 6151 80 39 100 E mail support hbm com www hbm com Modifications reserved All details describe our products in general form only They are not to be...

Отзывы: