30
TEDS
A1813-1.1 en
HBM
•
Click on the ”
Store and activate TEDS
” button to write the entered data
into the transducer TEDS memory and import them into the measuring
amplifier. Measuring can now take place, the TEDS Editor can be
terminated.
•
If necessary, implement zero adjustment in the MGCplus Assistant (view
”Signal processing”, column ”Zero value” or ”Tare value”; either enter a
value in the column Zero value/Tare value or zero using the displayed
button).
NOTE:
A zero adjustment is generally not stored in the TEDS memory as this is
dependent on the installation situation of the transducer. If, in special
individual cases, the zero adjustment must be stored in the TEDS memory,
an explicit value can be entered as a zero offset or tare.
This is implemented in the TEDS Editor on the ”Signal processing” tab.
This creates an additional HBM specific template which can then only be
interpreted by HBM measuring amplifiers. It is ignored by other measuring
amplifiers.