Calibration and Verification
Software Version 2.22

ISOBE5600

English

User Manual

I2716-4.0 en

Содержание Genesis ISOBE5600

Страница 1: ...Calibration and Verification Software Version 2 22 ISOBE5600 English User Manual I2716 4 0 en...

Страница 2: ...64293 Darmstadt Germany Tel 49 6151 80 30 Fax 49 6151 8039100 Email info hbm com www hbm com highspeed Copyright 2014 All rights reserved No part of the contents of this document may be reproduced or...

Страница 3: ...LICENSE AGREEMENT AND WARRANTY For information about LICENSE AGREEMENT AND WARRANTY refer to www hbm com terms ISOBE5600 I2716 4 0 en 3...

Страница 4: ...4 I2716 4 0 en...

Страница 5: ...5 4 Overvoltage current protection 22 2 5 5 Isolation 22 2 6 Environment 23 2 7 Laser Safety 24 2 8 Manual handling of loads 25 2 9 International safety warnings 26 2 10 Operation of electrical instal...

Страница 6: ...ibration 68 5 2 2 DC Calibration 72 5 2 3 AC Calibration depending on device 74 5 2 4 Calibration test results 79 Passed test results 80 Failed test results 80 6 Calibration Verification 82 6 1 Introd...

Страница 7: ...log file 110 B 2 1 The Extra log file 111 Verification test report 111 Test name DCgain test 112 B 3 Producing graphical results DC Gain 115 B 3 1 Producing graphical results Full Bandwidth 115 C Test...

Страница 8: ...death or serious injury CAUTION Indicates a potentially hazardous situation which if not avoided could result in minor or moderate injury or alerts against unsafe practices or alerts against actions...

Страница 9: ...d to Windows XP in Windows Vista and Windows 7 the Start Menu has undergone some significant changes The taskbar icon is no longer labeled Start and is now simply the pearl icon of the window frame in...

Страница 10: ...ver it is likely that safety protection has been impaired the instrument must be made inoperative and secured against any unintended operation Qualified maintenance or repair personnel should be infor...

Страница 11: ...n instructed to carry out installation commissioning maintenance or repair of the unit must have read and understood the User Manual and in particular the technical safety instructions Remaining dange...

Страница 12: ...of automation and test and measurement technology l Automation plant or test and measurement operating personnel Have been instructed how to handle the equipment and are familiar with the operation o...

Страница 13: ...ions manual may cause interference to radio communications As temporarily permitted by regulation it has not been tested for compliance with the limits for Class A computing devices pursuant to Subpar...

Страница 14: ...connected before the ground connection is removed i e the supply lead must be connected whenever signal leads are connected WARNING For safety it is essential that the protective ground connector of t...

Страница 15: ...WARNING If connection to a protective ground is not possible for any reason then please refer to the international safety standard EN 50191 2000 ISOBE5600 I2716 4 0 en 15...

Страница 16: ...chassis ground connection This point is not a protective ground Where caution is required this symbol refers to the user manual for further information This symbol warns that high voltages are presen...

Страница 17: ...be sustained by the equipment O Other previously referred to as CAT I This category is for measurements not directly connected to mains Think of measurement of signal levels regulated low voltage cir...

Страница 18: ...00 300 600 2 210 3 310 4 260 4 000 6 000 8 000 600 1 000 3 310 4 260 6 600 6 000 8 000 12 000 Using the above table one can deduct that this specification informs the user the device passed the insula...

Страница 19: ...5 s a c test V a c r m s Impulse test V peak Basic insulation and supplementary insulation Reinforced insulation Basic insulation and suplementary insulation Reinforced insulation 150 840 1390 1550 25...

Страница 20: ...gure 2 2 Figure 2 2 Acceptable arrangement of protective means against electric shock Example A measurement device is specified as 600 V CAT II reinforced insulation maximum input voltage 1000 V DC Us...

Страница 21: ...le damage or has been subjected to severe transport stresses It is the responsibility of the user to ensure the safety of any accessories such as probes used with the equipment WARNING ELECTRICAL SHOC...

Страница 22: ...st not be operated with the covers removed There are no user serviceable parts inside 2 5 5 Isolation CAUTION For input channels with plastic BNCs galvanically isolated from the chassis the input cond...

Страница 23: ...he ambient temperature If the instrument has a fan installed leave space around the equipment for unrestricted ventilation Do not store the equipment in hot areas High temperatures can shorten the lif...

Страница 24: ...set by government agencies for Class 1 laser products l In the USA the GEN series products are certified as a Class 1 laser product conforming to the requirements contained in the Department of Healt...

Страница 25: ...is load safely or is it a two person lift l How far will the load have to be carried l Is the path clear of clutter cords slippery areas overhangs stairs curbs or uneven surfaces l Will closed doors b...

Страница 26: ...ligvis blevet beskadiget og skal derfor kontrolleres af en kvalificeret tekniker Afbryd instrumentet fra lysnettet ved at fjerne IEC stikket Instrumentets vekselstr msafbryder er kun beregnet til funk...

Страница 27: ...ekwalificeerde en deskundige monteur te worden gecontroleerd Om dit instrument los te koppelen van de wisselstroomvoeding dient de IEC aansluiting er uit te worden getrokken De wisselstroom voedingssc...

Страница 28: ...stuttaa ammattihenkil ll Katkaise laitteen k ytt j nnite irrottamalla IEC liitin Laitteen verkkokytkimell on ainoastaan toiminnallinen tarkoitus Sit ei ole tarkoitettu eik se sovellu laitteen erottami...

Страница 29: ...n technicien qualifi Pour couper l alimentation secteur de cet instrument d brancher le cordon secteur L interrupteur d alimentation secteur sur cet instrument est purement fonctionnel Il ne s agit pa...

Страница 30: ...n von einem qualifizierten Elektriker gepr ft werden Zum Trennen des Ger tes von der Wechselstromversorgung den IEC Stecker abziehen Der Wechselstromversorgungs Schalter dient bei diesem Ger t nur f r...

Страница 31: ...ico specializzato e qualificato Per disinnestare questo strumento dall alimentazione a corrente alternata levare il connettore IEC L interruttore dell alimentazione a corrente alternata di questo stru...

Страница 32: ...en kvalifisert ingeni r For koble dette instrumentet fra AC tilf rselen trekker du ut IEC kontakten AC tilf rselsbryteren p dette instrumentet er kun for funksjonelle form l Den er ikke beregnet for e...

Страница 33: ...m qualifica o adequada Para desconectar este instrumento da alimenta o CA retire o conector IEC da ficha Neste instrumento o interruptor de alimenta o CA fornecido apenas para fins funcionais N o se d...

Страница 34: ...ser verificado por um engenheiro com qualifica o adequada Para desconectar este instrumento da alimenta o CA desconecte o conector IEC Neste instrumento o interruptor de alimenta o CA fornecido soment...

Страница 35: ...l t cnico especializado y autorizado al efecto Para desconectar este instrumento de la red desenchufe el conector IEC El interruptor de entrada de CA encendido se incluye solo para fines funcionales N...

Страница 36: ...r att kopplas loss detta instrument fr n str mf rs rjningen dra ut IEC kontakten Brytaren f r v xelstr mf rs rjningen p detta instrument r endast avsedd f r funktionella syften Den r inte avsedd eller...

Страница 37: ...cked by a suitably qualified engineer To disconnect this instrument from the AC supply unplug the IEC connector The AC supply switch on this instrument is provided for functional purposes only It is n...

Страница 38: ...33V RMS 46 7V RMS 70V DC IEC 61010 1 2010 AC AC IEC AC EN 50110 1 EN 50110 2 50 V AC RMS 120 V DC ISOBE5600 38 I2716 4 0 en...

Страница 39: ...33 V RMS 46 7 V 70 V DC IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 V AC RMS 120 V DC ISOBE5600 I2716 4 0 en 39...

Страница 40: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 ISOBE5600 40 I2716 4 0 en...

Страница 41: ...33 V RMS 46 7 V 70 V DC IEC 61010 1 2010 AC AC IEC AC EN 50110 1 EN 50110 2 50 V AC RMS 120 V DC ISOBE5600 I2716 4 0 en 41...

Страница 42: ...maintenance procedures The European Standard EN 50110 2 is a set of normative annexes one per country which specify either the present safety requirements or give the national supplements to these mi...

Страница 43: ...ng over a carpet while scuffing your feet building up electrons on your body you may get a shock the discharge event when you touch a metal doorknob This little shock discharges the built up static el...

Страница 44: ...assis at the other WARNING The wrist strap is intended for static control only It will not reduce or increase your risk of receiving an electric shock from electrical equipment Follow the same precaut...

Страница 45: ...user is encouraged to try to reduce the interference by one or more of the following measures l Re orient or relocate the affected equipment l Increase the distance between the instrument and the affe...

Страница 46: ...ations old devices that can no longer be used must be disposed of separately and not with normal household garbage If you need more information about waste disposal please contact your local authoriti...

Страница 47: ...3 3 Declaration of conformity For information about the CE Declaration refer to www hbm com highspeed ISOBE5600 I2716 4 0 en 47...

Страница 48: ...and the medium voltage isolated transmitter model ISOBE5600tm The ISOBE5600 Calibration and Verification software tests against the published specifications Regardless of whether any remedy set forth...

Страница 49: ...be able to perform all the required steps several parts are required Order No Article 1 GN CAL8 2 ISOBE5600 Calibration and Verification Software 1 GN CAL CABLES 2 GEN DAQ Calibration basic cable set...

Страница 50: ...Minimum RAM memory 1 GByte Minimum free disk size 200 MB Minimum graphics card 16 bit color with 64 MB on board video memory and hardware DirectX 9 support Minimum screen resolution 1024 x 768 pixels...

Страница 51: ...4 4 Supported acquisition hardware Supported ISOBE5600 receivers l ISOBE5600r l ISOBE5600m Supported ISOBE5600 transmitters l ISOBE5600t l ISOBE5600tm ISOBE5600 I2716 4 0 en 51...

Страница 52: ...e 5720A Calibrator LF Generator l Fluke 5820A Fluke 5900B Oscilloscope Calibrator HF Generator l Hewlett Packard HP 3458A Multimeter DVM Required calibrators Fluke 5700A or Fluke 5720A Fluke 5820A or...

Страница 53: ...ill fail to connect to the GPIB controller Note The most current version of the software at release of this manual still supports the legacy IOTech serial to GPIB interface Connecting the hardware For...

Страница 54: ...Figure 4 1 Connecting directly to your PC with the NI GPIB USB H USB controller ISOBE5600 54 I2716 4 0 en...

Страница 55: ...the ISOBE5600 Calibration and Verification AutoPlay dialogue box If the AutoPlay dialogue box does not appear choose Start Run enter d setup exe where d is the letter of your CD ROM drive and click O...

Страница 56: ...and Verification software CAUTION Before starting a calibration or verification it is recommended that the ISOBE5600 system the receiver transmitters and the calibration standards run at least one ho...

Страница 57: ...The ISOBE5600 Calibration and Verification window will be displayed Figure 4 3 ISOBE5600 Calibration and Verification window ISOBE5600 I2716 4 0 en 57...

Страница 58: ...ion then is stored within the PNRF file for automatic traceability of your calibration information Enter Organization information When you start the ISOBE5600 Calibration and Verification software the...

Страница 59: ...u want to modify your organization information Figure 4 5 Change Organization information A Modify 1 Click Modify 2 Make the required changes Figure 4 6 Organization changes 3 Then click Save to save...

Страница 60: ...alibration and Verification window A Mainframe list In the Mainframe list click the ISOBE5600 mainframe that you want to calibrate or verify The ISOBE5600 serial number uniquely identifies each ISOBE5...

Страница 61: ...After a mainframe has been selected the Mainframe Selection window will be displayed Figure 4 8 Mainframe Selection window ISOBE5600 I2716 4 0 en 61...

Страница 62: ...nsmitter Receiver selection A Transmitter Receiver check box B Channel list In the Transmitter Receiver check box check Transmitter or Receiver In the Channel list click a channel to display informati...

Страница 63: ...ows which calibration standard s you are using In the Mainframe Selection window Figure 4 10 Mainframe Selection window A Calibrate B Verify 1 Click Calibrate to open the Board Calibration window or c...

Страница 64: ...Figure 4 11 Board Calibration window Transmitter Figure 4 12 Board Verification window Receiver A Set up CalStack ISOBE5600 64 I2716 4 0 en...

Страница 65: ...t component And if you are using the IOtech Serial IEEE488 l The COM port of the IOtech Serial IEEE488 controller matches the settings of your IOtech configuration For details please refer to the IOte...

Страница 66: ...tware will return to the Calibration or Verification window 4 8 6 Exit the ISOBE5600 Calibration and Verification software You can exit the ISOBE5600 Calibration and Verification software in the ISOBE...

Страница 67: ...ndards run at least one hour to achieve the best test specifications CAUTION HBM uses state of the art electronic components in its equipment These electronic components can be damaged by discharge of...

Страница 68: ...alibrate the Transmitter or the Receiver 5 2 1 Starting the calibration In the Mainframe Selection window Figure 5 1 Mainframe Selection window A Calibrate Select the probe system you wish to calibrat...

Страница 69: ...ows information about the selected device C Start To start the calibration process D Pause To pause or stop while testing E Close To end the calibration process F Set up CalStack To open the CalStack...

Страница 70: ...tion or verification to ensure that the ISOBE5600 Calibration and Verification software knows which calibration standard s you are using To begin the calibration process of the selected device 1 Click...

Страница 71: ...ected click OK to start testing The system will start the calibration process Figure 5 4 DC Calibration Transmitter A Test Indicates the test to be performed B Active test progress bar ISOBE5600 I2716...

Страница 72: ...e DC Calibration B When the test is passed the background behind DC Calibration will turn to green C When the test is failed the background behind DC Calibration will turn to red Note If you performed...

Страница 73: ...The Active test progress bar gives information on what is currently happening Figure 5 7 Active test progress bar ISOBE5600 I2716 4 0 en 73...

Страница 74: ...ment When the DC Calibration test is finished the calibration process continues with the AC Calibration test 1 Connect the testing fixture to the calibration standard s as displayed on your screen Not...

Страница 75: ...r other non conductive adjustment tool CAUTION HBM uses state of the art electronic components in its equipment These electronic components can be damaged by discharge of static electricity ESD Theref...

Страница 76: ...sition of the adjustments is depends on the device Please refer to appendix A ISOBE5600 Calibration Adjustments on page 119 of this manual Figure 5 9 AC Calibration adjustments example A T118 B T127 N...

Страница 77: ...edle enters the green section Repeat this adjustment for each channel When all channels have been properly adjusted the Next Step or Ready button depending on the selected device will be enabled Figur...

Страница 78: ...y T118 Click Ready after you have adjusted all channels The software will then continue the calibration process Note If you do not adjust a particular channel or if you click Skip Channel s the calibr...

Страница 79: ...alibration tests is failed l A Calibration failed message appears l The failed tests and results are indicated by a red background You can see the test results once all tests have been performed Figur...

Страница 80: ...a Saving Calibration date message appear in the bottom left hand corner of the screen Figure 5 13 Saving Calibration date message Failed test results If one of the tests is failed a Calibration failed...

Страница 81: ...d background The ISOBE5600 Receiver and Transmitter are calibrated independently Transmitter calibration does not depend on the channel number The calibrated Transmitter can be verified in each channe...

Страница 82: ...ion Ratio l DC Output l Output Noise l Output Res Note The calibration and verification tests depend on the installed and selected board Therefore the windows displayed may differ from the illustratio...

Страница 83: ...p the Verification test In the Mainframe Selection window Figure 6 1 Mainframe Selection window A Verify Click Verify to open the Board Verification window The Board Verification window will be displa...

Страница 84: ...n process F Save Results To save the verification test results G Pause To pause or stop while testing H Close To end the verification process I Set up CalStack To open the CalStack Setup window J Test...

Страница 85: ...e only the 3 db bandwidth value Figure 6 3 Board Verification window Bandwith tests A Full Bandwith Selecting the Full check box This test is longer than the quick bandwidth test It tests more input v...

Страница 86: ...device 1 Click Start 2 Connect the testing fixture to the calibration standard s as displayed on screen Figure 6 4 Example of an ISOBE5600 Connection window Note The color of your cables may differ f...

Страница 87: ...6 2 3 During Verification Testing Figure 6 5 DC Gain A Test Indicates the test to be performed B Active test progress bar C Test results ISOBE5600 I2716 4 0 en 87...

Страница 88: ...ain b When the test is passed the background behind DC Gain will turn to green c When the test is failed the background behind DC Gain will turn to red If the receiver verification was chosen you will...

Страница 89: ...test progress bar gives information on what is currently happening Figure 6 8 Active test progress bar C Test results For more information please refer to chapter Verification test results on page 90...

Страница 90: ...he verification tests is failed l A Verification failed message appears l The failed tests and results are indicated by a red background You can see the test results once all tests has been performed...

Страница 91: ...ate message Failed test results Failed tests and results are indicated by a RED background If one of the tests is failed a Verification failed message appears Figure 6 11 Verification failed message N...

Страница 92: ...lts to save the verification test results to a report The report is a Word file containing l Mainframe information l Board recorder information l Verification test results for each test on each channe...

Страница 93: ...Mode Filter and Range settings supported by the Device Under Test DUT The selection is done based on the technical design of the DUT to ensure all design combinations are tested Procedure Six voltage...

Страница 94: ...mum Absolute Vmeasi Vini Range 100 7 1 1 The MIL Military standard test This is also a DC gain test and is different to the basic test in several ways It covers all voltage ranges and all filters and...

Страница 95: ...ks in all possible settings combinations Procedure l A DC signal 90 full scale is applied l 4000 samples are acquired l The standard deviation of this data is calculated l This value must be 0 5 signa...

Страница 96: ...ne the 3 dB point If Full bandwidth is selected an enhanced test is also performed on the different signal frequencies 7 3 1 Quick Bandwidth Test AC gain is measured using a total of 3 frequency value...

Страница 97: ...plifier supports The start frequency is set to maximum sample rate of the channel 2000 Fstart The end frequency is set to the maximum sample rate of the channel Fstop The frequency range from Fstart t...

Страница 98: ...annel Vmeasured Note Actual frequencies used during the tests are rounded to the step size of the calibrator used and are slightly increased to overcome beat frequency measurements The results logged...

Страница 99: ...s HINT TIP The noise test uses a selected set of all Mode Filter and Range settings supported by the Device Under Test DUT Typically in one filter setting all ranges are tested where all other filter...

Страница 100: ...from the LF generator is between channel input signal and isolated ground and system ground For non isolated channels only differential amplifiers can be CMRR tested and then the AC signal is applied...

Страница 101: ...voltages are generated by the ISOBE5600 Receiver one at a time The voltages are equally divided from 45 to 45 of Range and are denoted as Vouti These output voltages are measured with the Multimeter...

Страница 102: ...timeter Isolated Probe System settings l The Output Noise test can be done without the Transmitter connected Procedure The output is set to 0 V The output AC voltage rms is measured with the Multimete...

Страница 103: ...test can be done without the Transmitter connected Procedure The output is set to 90 full scale This output voltage is measured with the Multimeter twice First without resistive load DC reference Rloa...

Страница 104: ...set to calibrate and verify the ISOBE5600 systems Using this calibration kit the proper calibration equipment and calibration fixtures enable easy fast and on site self calibration of the ISOBE5600 sy...

Страница 105: ...BE5600r and ISOBE5600m Transmitters ISOBE5600t and ISOBE5600tm Calibration Equipment Not supplied by HBM Beyond the calibration kit itself the following calibration equipment is needed in order to per...

Страница 106: ...s Exact process description and direct readouts make even manual procedures easy to work with Figure A 3 ISOBE5600 AC adjustment description and readouts Verification Process During the fully automate...

Страница 107: ...Calibration Report Printout As an end result the ISOBE5600 system saves all the results in a RTF Text file From there it can be stored for later reference or printed out ISOBE5600 I2716 4 0 en 107...

Страница 108: ...ns to be ordered separately 1 Article Description Order No GEN series Calibration basic cable set GEN series Calibration basic cable set contains cables adapters and termination resistors to ensure a...

Страница 109: ...cations file Standard verification report rich text Standard verification report plain text Extra report Standard verification Yes Yes Yes Yes DC gain test measured values Full bandwidth verification...

Страница 110: ...Physical Name ISOBE5600 Serial Number IES0700097 Type ISOBE5600 SW Version 1 10 7339 No of channels 4 Channel Info Physical Name Ch 4 Serial Number IET0800103 Channel Type Fiber Amplifier Rv1 Channel...

Страница 111: ...016 NA 20 0 0 001 0 016 0 005 0 009 NA NA 0 016 NA 40 0 0 000 0 002 0 005 0 004 NA 113 1 0 018 NA 100 0 0 002 0 026 0 005 0 013 NA NA 0 016 NA Note This table shows only a part of the report B 2 1 Th...

Страница 112: ...1 Ch 2 Ch 3 Ch 4 0 180000 0 180040 0 000000 0 000000 0 000000 0 108000 0 108046 0 000000 0 000000 0 000000 0 036000 0 036027 0 000000 0 000000 0 000000 0 036000 0 035968 0 000000 0 000000 0 000000 0 1...

Страница 113: ...Vpp Freq Chn 1 Hz dB 60000 0 050 100000 0 060 160000 0 056 250000 0 059 ni 400000 0 058 630000 0 062 1000000 0 056 1590000 0 051 2510000 0 042 3990000 0 016 6320000 0 017 10010000 0 112 15860000 0 52...

Страница 114: ...OM OTHER RANGES FOR DOCUMENATION PURPOSES Test name Bandwith test Mode Positive Range 2 V Filter Bessel_AA Test amplitude 0 18 Vpp Freq Chn 1 Hz dB 60000 0 075 TEXT REMOVED FROM OTHER RANGES FOR DOCUM...

Страница 115: ...gure B 1 Example of a graphical output DC Gain Test One extra column highlighted in blue has been created to show the difference between the two columns of the expected and actual results Doing this w...

Страница 116: ...type Delimited Delimiter Semi colon to be able to import the data correctly Figure B 2 is an example of the expected graphical output for a selected Full bandwidth test Figure B 2 Example of a graphic...

Страница 117: ...00 calibration kit is the basic software and tools required to get started 1 GN CAL8 2 ISOBE5600 Calibration kit Part number Number Description LD 426 036200 1 Software CD inc Manual V1 10 LD 869 9118...

Страница 118: ...LD 085 997900 2 CBL BNC SAFETY LD 024 924700 1 CON INSULATED SHORT LD 085 998100 1 BANANA CABLE 1 50 M RED ISO LD 085 009000 1 BANANA CABLE 1 50 M RED ISO modified LD 085 998300 1 BANANA CABLE 1 50 M...

Страница 119: ...ross head fasteners and plugs at each end of the Isolated Digitizer outer casing 5 Remove the blue casing lid to reveal the metal battery housing 6 Carefully lift out the metal battery power supply ca...

Страница 120: ...1 Accessing the Isolated Digitizers A Handle of battery Only when working with the ISOBE5600t B Battery power supply casing C CBL 6600 HV ID AC CAL PWR D Isolated Digitizer Casing ISOBE5600 120 I2716...

Страница 121: ...ent points A T118 Adjustment point DIV 10 B T127 Adjustment point DIV 100 Closing the Isolated Digitizers When AC calibration is completed reassemble the Isolated Digitizer in reverse order from the o...

Страница 122: ...commended to replace the IOTech with a USB to GPIB convertor l Connect the calibrator equipment to the GPIB side of the IOTech serial to GPIB interface l Connect the serial side of the IOTech interfac...

Страница 123: ...st 84 C Calibrate 63 68 Calibration test results 79 Failed test results 80 Passed test results 80 Calibrator 52 CalStack setup 63 69 84 Cancel 65 Categories according to IEC 61010 2 030 2010 17 Cautio...

Страница 124: ...eady 78 Receiver DC Out Cal 67 Report 92 Required fixtures 49 S Safety Environment 23 Fuse 21 Instrument symbols 16 International safety warnings 26 Manual handling of loads 25 Safety Messages Appropr...

Страница 125: ...Transmitter DC Calibration 67 Transmitter Receiver 62 Channel information 62 V Verification test report 111 Verification test results 90 Failed test results 91 Passed test results 91 Report 92 Save re...

Страница 126: ...126 I2716 4 0 en...