HI 6600 Series Modular Sensor System User Guide
Page | 125
4: WRITE NONVOL CMD
. Write a #4 to the command register to save parameters in non-volatile
memory.
o
No Error Codes
5: CMD
: Reserved
o
No Error Codes
6: WEIGHT SAMPLE CMD:
Write a #6 to the command register to run the Weigh
Sample command to set up the calibration of the Counts (if enabled).
o
Status Error code 1 (motion)
o
Status error code 2 (A/D error)
•
0x64 (100 decimal): CAL LOW CMD
. Write a 0x64 hex to the command register to perform the
low step of a traditional calibration. The status register will read 0 if this command succeeds:
o
Calibration_Fail 1
o
Calibration_Fail_Motion 3
o
Calibration_Fail_Adc_Error 4
•
0x65 (101decimal): CAL HIGH CMD
. Write a 0x65 hex to the command register to perform the
high step of a traditional calibration.
o
Calibration_Fail 1
o
Calibration_Fail_Motion 3
o
Calibration_Fail_Adc_Error 4
o
HardcalFailCounts 8: not enough counts between hard cal hi and hard cal lo
•
0x66 (102 decimal): C2 CAL CMD
. Write a 0x66 hex to the command register to perform a C2
calibration.
o
Calibration_Fail 1
o
Calibration_Fail_Motion 3
o
Calibration_Fail_Adc_Error 4
o
Calibration_Fail_Noc2 5
o
Calibration_Fail_C2capeq 6
o
Calibration_Fail_C2clones 7
•
0x1000 (4096 decimal): WRITE INTEGER CMD.
Set the value of an integer parameter.
Write 0x1000 in the command register 0, the parameter ID number in registers 2, 3 and the
desired value in registers 4, 5.
o
No Error Codes
•
0x1001 (4097 decimal): WRITE FLOAT CMD.
Set the value of a floating point parameter.
Write 0x1001 in the command register 0, the parameter ID number in registers 2, 3 and the
desired float value in registers 4, 5.
•
Failure 1
•
TooHigh -1 : value is above legal limit
•
TooLow -2 : value is below legal limit
•
ErrorNotFound 0x8000 : parameter id not found
Содержание HI 6600-EIP
Страница 16: ...HI 6600 Series Modular Sensor System User Guide Page 16...
Страница 44: ...HI 6600 Series Modular Sensor System User Guide Page 44...
Страница 96: ...HI 6600 Series Modular Sensor System User Guide Page 96 General Troubleshooting Flow Chart Index...
Страница 97: ...HI 6600 Series Modular Sensor System User Guide Page 97 A ElectricalandMechanicalReview...
Страница 98: ...HI 6600 Series Modular Sensor System User Guide Page 98 A1 CheckingforUnstableComponentsinaWeighingSystem...
Страница 100: ...HI 6600 Series Modular Sensor System User Guide Page 100 B1 GuidelinestoVerifyElectricalInstallation...
Страница 101: ...HI 6600 Series Modular Sensor System User Guide Page 101 B2 GuidelinestoVerifyMechanicalInstallation...
Страница 102: ...HI 6600 Series Modular Sensor System User Guide Page 102 B3 VerifyConfiguration FilterSettingstoImproveStability...
Страница 103: ...HI 6600 Series Modular Sensor System User Guide Page 103 C IntegratedTechnicianandStabilityTestOverview...
Страница 104: ...HI 6600 Series Modular Sensor System User Guide Page 104 ENon ReturntoZero SystemwithITSummingCard...
Страница 106: ...HI 6600 Series Modular Sensor System User Guide Page 106 G Calibration Errors During Calibration...
Страница 107: ...HI 6600 Series Modular Sensor System User Guide Page 107 H Mechanical Installation...
Страница 108: ...HI 6600 Series Modular Sensor System User Guide Page 108 J Electrical Inspection...
Страница 109: ...HI 6600 Series Modular Sensor System User Guide Page 109 K InstallationCheckPoints...
Страница 110: ...HI 6600 Series Modular Sensor System User Guide Page 110 M WeightProcessor sOptionalFrontDisplayBlankorLocked...
Страница 128: ...HI 6600 Series Modular Sensor System User Guide Page 128...
Страница 130: ...HI 6600 Series Modular Sensor System User Guide Page 130 Image of HI 6600 Hardy Gateway Module II Diagram...
Страница 131: ...HI 6600 Series Modular Sensor System User Guide Page 131 Image of HI 6610 Weight Processing Module II Diagram...
Страница 132: ...HI 6600 Series Modular Sensor System User Guide Page 132 OptionalHI6110DisplayPanelMountingTemplate...