48
MENU
Value
This parameter defines the Set point value. With Value selected, press
Set
. Press the
keys to
edit the required value within minimum / maximum probe limits (e.g. 0.00 to 12.00 pH), displayed
blinking. Press
CFM
, to save.
Mode
Option:
ON/OFF, Proportional, PID
The Mode parameter defines the type of control the controller will use: ON/OFF, Proportional or PID.
The
Setup
virtual key is used after selecting
Mode
, to set additional settings. See SECTION 14 for
detailed information on Control Modes & Algorithms.
Navigation
• To open the drop-down list for Mode, press
Modify
.
• Press the
keys to select mode type: ON/OFF, Proportional or PID.
• Press
Select,
to save.
• After selecting
Mode
, press
Setup
.
Setup for ON/OFF control
• For the drop-down list to be displayed, press
Setup
.
• Press the
keys to move between Mode
and
Hysteresis.
• Select Mode and press the virtual keys to choose ON/OFF Low or ON/OFF High.
• Press the key, to select Hysteresis.
• With Hysteresis (default value 1.00 pH) highlighted, press
Set.
The present value will blink
permitting editing, within minimum 0.02 pH and maximum 1.80 pH, using the
keys.
• Press
CFM,
to save.
• Press the key, to exit Setup.