115
CLEANING MODE
• At a transition to manual mode. Cleaning cycle is stopped instantly. After exiting from manual
mode, cleaning will continue with a rinse and a recovery phase.
Time
Time
Time
Time
Cleaning interval
Cleaning
cycle
Cleaning
cycle
Cleaning
cycle
Cleaning
cycle
Cleaning
cycle
Cleaning
cycle
Cleaning triggered by internal Timer
Internal Timer
Stop condition
Cleaning
Cleaning cycle on
stop condition
Figure 54:
End Cleaning, Stop Condition
15.6. OVERVIEW OF CLEANING TYPES
Cleaning
Triggers
Associated Relay
Steps
Simple
(water only)
• internal triggers
(timer
¹,
schedule²)
• external trigger
• manual start
any non-assigned
relay can be set
as Rinse Relay
• instrument enters HOLD mode
• configured relay(s) is (are) energized
Advanced
(water &
detergent)
• internal triggers
(timer
¹
, schedule²)
• external trigger
• manual start
any non-assigned
relay can be set
for advanced
cleaning (i.e. at
least one Rinse
Relay and Wash
Relay)
• instrument enters HOLD mode
• rinse is energized (pre-rinsing
phase)
• wash is energized (detergent phase)
• wash is de-energized (post-rinsing
phase)
¹ If the trigger is set to timer, entering in a Hold mode that overlaps with the trigger will add a delay to
the cleaning cycle.
² If the trigger is set to schedule, entering in a Hold mode that overlaps with the trigger of the next start
time and exceeds the set time, the trigger will be lost. The trigger will still be activated if the internal clock
doesn't get past a minute over start time.