10. GENERAL OPERATIONS
10.1. POWERING THE CONTROLLER
Measure is the normal operating mode of the controller. Depending on the probe wired, and the
location of the probe, the display will show measured values with appropriate measurement units.
At start-up, while the controller performs internal checks, the display will show the Hanna Instruments
logo, controller name, date, and firmware version.
In Measure mode, with the probe connected, the controller detects the probe and the probe type.
With no probe connected or a new probe connected, at start-up the controller can display one of the
warning messages detailed below.
Warning Message
Description
“No probe connected”
• No probe has been connected
• A connection issue has been detected
“Different probe. Please set control
parameters.”
A different probe type (different probe series) has
been connected.
“New probe. Update control settings if
necessary.”
A new probe (same series) has been connected.
A “Startup delay” message, associated with a programmable countdown timer indication, is displayed
at power on.