19
Recurrent Performance and Safety Testing
The following table highlights AP-450
’s Essential Performance Parameters, with their subsequent testing
method and frequency. If your AP-450 does not meet the following criteria, contact your local Handicare
authorized dealer for service.
Performance Criteria
Recommended Check
Frequency
The AP-450 should be
capable of lifting 450 lb
(204 kg).
Load test at full load
(450lb/204kg)
Annual
The AP450 will operate as
intended throughout its
service life, provided all
recommended
maintenance and usage
recommendations are
followed.
1. Load test at full load
(450lb/204kg)
2. Pre-lifting inspection by
user care giver.
3. Preventive maintenance
by trained tech.
1. Annual.
2. Every use
3. Semi-annually or annually, per usage
frequency.
“General Inspection and Preventive
” on page 17.
The AP450 should be
capable of working in
conjunction and
compatibility with
Prism/Handicare Track,
trollies and accessories
(Carry bars and Slings).
1. Load test at full load (450
lb)
2. Pre-lifting inspection by
user care giver.
3. Preventive maintenance
by trained tech.
1. Annual.
2. Every use
3. Semi-annually or annually, per usage
frequency.
See “General Inspection and Preventive
Maintenance
” on page 17.
The AP450 should have
“Free Fall” protection
system.
Preventive maintenance by
trained tech.
Semi-annually or annually, per usage
frequency. See
The AP-450 should have
upper and lower limits.
Preventive maintenance by
trained tech.
Semi-annually or annually, per usage
frequency. S
The AP-450 strap should
have a minimum safety
factor of 6 in terms of
break load.
1. Load test at full load (450
lb)
2. Pre-lifting inspection by
user care giver.
3. Preventive maintenance
by trained tech.
Annual
Every use
Semi-annually or annually, per usage
frequency
See “General Inspection and Preventive
Maintenance
The AP-450 should have
visual “Error” status
display.
Preventive maintenance by
trained tech.
Semi-annually or annually, per usage
frequency.
See “General Inspection and Preventive
Maintenance
The AP-450 hand control
locking mechanism used
as connections should be
“Self-Locking”.
Preventive maintenance by
trained tech.
Semi-annually or annually, per usage
frequency.