17
Fault
Recommended Solution
One side of the lift strap is
starting to fray after
continued use.
Discontinue using immediately and replace strap. Check to be sure
that the lift is always directly above the individual being lifted,
especially with power traversing lifts. Refer to
” on page 11 for correct lift positioning. If fraying still
continues, contact your local authorized dealer immediately to check
the lift and ensure proper continued operation.
The lift does not pass
through a track
component such as a
turntable or transistion
gate.
Refer to the Owner
’s Manual for the specific piece of equipment in
question. If the recommended solution does not correct the problem,
contact your local authorized dealer immediately to check the lift and
ensure proper continued operation.
No Power
Ensure the Emergency Stop Switch is set to ON.
Lift starts beeping while
pushing Emergency Down
button with lift not coming
down
The AP-450 lift is equipped with dual sensors for strap tension and
position, the Emergency Down Button/Switch system only works
with load on the strap. Emergency Down Button/Switch motion stops
immediately after the lift strap is loose.
The lift is recommending
preventive maintenance.
Complete the “Semi-Annually” maintenance procedure on page 18
prior to resetting the lift counter. Instructions to reset the lift counter
are detailed in the AP-450 Service/Spare Parts Manual.
General Inspection and Preventive
Maintenance
Do not perform any inspection or maintenance while the lift is in use with a
patient.
Each Use
(To be completed by User)
Prior to each use the AP-450 lift and associated track, accessories, and slings must be visually
inspected. Refer to the accessory and sling user guides for specific details regarding their
inspection.
Should any of these items fail the inspection do not use the lift. Contact your
local authorized dealer for service.
Visually check for the following:
The lift strap shows NO signs of fraying or breaking along its entire length.
The stitching on the lift strap where it connects to the carabiner shows no signs of fraying
or breaking.
If the lift strap wear indicators are grey, replace the lift strap regardless of number of
cycles.
The sling that will be used show no signs of unusual wear and tear. The straps of the sling
that connect to the carry bar of the lift show no signs of fraying or breaking. Refer to
specific sling instructions.
The Hand Control cable is not kinked, twisted, knotted, disconnected, or damaged.
All the functions on the Hand Control work correctly (e.g. Up/Down, etc.).
There are no cuts, dents or sharp edges on the hooks that may damage the sling straps.
The lift has no unusual sounds when it is moved up, down, left or right.
Ensure that end stops are installed at each end of the track.