11
5-3 Side
panels
①
Cooling fan air intake vents Air intake vents for cooling the internal power supply. When
installing this unit, provide a space of at least 10 cm from these
vents to a wall or other objects.
②
Air exhaust vents
Air exhaust vents for cooling the internal power supply. When
installing this unit, provide a space of at least 10 cm from these
vents to a wall or other objects.
③
Gas inlet port
Gas can be introduced into the sample chamber by inserting an
air tube (outer diameter: 6 mm) into this port. To introduce gas
into the sample chamber, close the ozone outlet and air vents in
the sample chamber.
(NOTE)
Even in that case, always connect a local exhaust ventilation
system or an ozone decomposition unit E12685 (sold
separately) to the duct connection port of this unit.
Never use flammable gases.
NOTE: See section 5-4 “Sample chamber” for details.
③
①
②