background image

19 

 

8. 

Maintenance and inspection 

To use this product correctly and safely, perform periodic maintenance and inspections. 
 

  Carefully read this manual and comply with the instructions to ensure safe use. 

  If a failure is suspected, please read the "Troubleshooting" section. 

  If this product gets wet or if oil or dust gets inside, the electrical insulation will deteriorate leading 

to hazardous situations such as electrical shock or fire. Deteriorated insulation may cause the 
lamp to malfunction or fail to light up. If this product becomes wet or contaminated with oil or dust, 
then immediately stop using it and contact our sales office for advice. 

  To clean the body of this product, gently wipe with soft, dry cloth. Do not use chemicals such as 

benzine and thinner. 

  If the lamp becomes dirty, turn off the lamp and also the POWER switch. Then wait until the lamp 

temperature drops sufficiently and then wipe the lamp with a clean, soft cloth moistened with 
alcohol. 

  The lamp is subject to wear. Replace the lamp with a new one before the total lamp ON time 

reaches 1000 hours. The lamp output intensity will drop or the lamp may be damaged after a 
long-term operation. 

  If the lamp is damaged, please return it to us without attempting any repair, etc. 

 
 

9. Troubleshooting 

 
 
 

Symptom Cause 

Corrective 

action 

Power will not turn on. 

Power cable is 
disconnected. 

Check that the power cable is securely connected. 

Lamp will not light up. 

Total lamp ON time has 
reached 1000 hours. 

The lamp must be replaced. Please contact us. 

Interlock is activated. 

Check that the sample chamber is completely 
closed. 

Irradiance is low. 

Dust or dirt is on the 
lamp. 

Remove dust and dirt on the lamp by wiping with 
gauze moistened with alcohol.   

“FAN2 ERR” is 
displayed. 

Shorting connector for 
external fan or connector 
of ozone decomposition 
unit (sold separately) is 
not connected to the 
EXT. FAN connector. 

Check that the supplied shorting connector for 
external fan or the connector of the ozone 
decomposition unit (sold separately) is securely 
connected to the EXT. FAN connector on the rear of 
this unit. 

“ERROR: LAMP ERR, 
FAN1 ERR, 
TEMPERATURE, or 
POWER SUPPLY” is 
displayed. 

Lamp, fan, or power 
supply has a problem. 

Contact us. 

Содержание EX-mini L12530

Страница 1: ...he product for easy reference when needed If this manual is lost or damaged promptly order a replacement copy from our sales office or sales representative If transferring this product to another part...

Страница 2: ...s inlet port 13 7 How to use 14 7 1 Pre operation check 14 7 2 Operation 14 7 3 Turning off the unit 15 7 4 Replacing the lamp 15 7 5 External control 16 7 6 Alarms and errors 18 7 7 Precautions durin...

Страница 3: ...the equipment This symbol indicates a precaution you must take Read the instructions carefully to ensure correct and safe product use This symbol indicates a prohibited action Read the instructions ca...

Страница 4: ...ar safety glasses gloves and other appropriate protective gear when operating this unit Refer to JIS T8141 or equivalent standards If the lamp bursts the glass fragments may cause injury When light is...

Страница 5: ...y after operation The section around the lamp is very hot during operation or immediately after operation Touching it may cause burns If the lamp is cracked stop using it Operating a cracked lamp may...

Страница 6: ...filled with xenon gas When disposing of this product take appropriate measures in compliance with applicable regulations regarding waste disposal and correctly dispose of it yourself or entrust prope...

Страница 7: ...e clearly visible at all times If the label peels off or becomes dirty replace it with a new label Please contact our sales office for a replacement warning label available for a charge Top view High...

Страница 8: ...it E12685 sold separately External dimensions 344 mm W 230 mm D 233 mm H not including projecting parts Sample chamber dimensions 204 W 139 D 118 H not including projecting parts Weight Approx 7 0 kg...

Страница 9: ...for irradiation time setting MODE button Use to select irradiation mode Button is lit in AUTO mode ON OFF button Use to turn the lamp on or off Button is lit when the lamp is lit POWER switch Pressin...

Страница 10: ...e decomposition unit always plug the supplied shorting connector for external fan into this connector EXT CONTROL external control connector When using an external control unit connect it to this conn...

Страница 11: ...f at least 10 cm from these vents to a wall or other objects Gas inlet port Gas can be introduced into the sample chamber by inserting an air tube outer diameter 6 mm into this port To introduce gas i...

Страница 12: ...he ozone outlet Vent cover When introducing external gas into the sample chamber use this cover to close the air vents Gas inlet port This port is for introducing external gas into the sample chamber...

Страница 13: ...ds that come with the E12685 CAUTION Before connecting the power cable make sure that the POWER switch on this unit is OFF 6 2 Connecting the power cable Connect the supplied power cable to the AC INP...

Страница 14: ...2 MANUAL mode In MANUAL mode you can manually turn the lamp on and off without setting the irradiation time Press the LAMP ON OFF button to light up the lamp the LAMP ON OFF button also lights up when...

Страница 15: ...mp is ON After the lamp lights up the set irradiation time on the display panel starts counting down The lamp automatically turns off when the irradiation time on the display panel reaches 0 zero or t...

Страница 16: ...lamp lights up by shorting pins 3 and 4 for 100 milliseconds or more When the lamp is lit in AUTO mode shorting pins 3 and 4 for 100 milliseconds or more turns off the lamp 4 Auto irradiation 5 Lamp...

Страница 17: ...50 mA or less Shorting the signal output terminal to the GND terminal outputs a signal Control signal input The ON contact point and transistor rating should be as follows Withstand voltage 30 V or mo...

Страница 18: ...ERROR POWER SUPPLY A power supply error occurs if a power supply failure occurs The ERROR POWER SUPPLY message appears on the display panel 7 6 6 Temperature error ERROR TEMPERATURE A temperature err...

Страница 19: ...eplace the lamp with a new one before the total lamp ON time reaches 1000 hours The lamp output intensity will drop or the lamp may be damaged after a long term operation If the lamp is damaged please...

Страница 20: ...r man made disasters 4 Failure or trouble was caused by malfunctions of another piece of equipment connected or used near this unit 5 Manufacturer warranty period has expired for components with a spe...

Страница 21: ...21 11 Dimensional outlines Unit mm...

Страница 22: ...l lokoldbaek hamamatsu de The Netherlands Office PO Box 50 075 NL 1305 AB Almere Netherlands Telephone 31 36 5382 123 Fax 31 36 5382 124 E mail info hamamatsu nl Poland Office 02 525 Warszawa ul sw A...

Отзывы: