16
7-5 External
control
This unit can be controlled by external control. Use the supplied “external control connector” to
control this unit by external control.
Table 1: Description of external control connector XM2F-1540 (made by OMRON)
Pin No.
Signal name
IN/OUT
Description
1
Manual irradiation (+)
IN
The lamp lights up as long as pins 1 and 2 are
shorted. (manual irradiation)
2
Manual irradiation (-)
-
3
Auto irradiation (+)
IN
The lamp lights up by shorting pins 3 and 4 for 100
milliseconds or more. When the lamp is lit in AUTO
mode, shorting pins 3 and 4 for 100 milliseconds or
more turns off the lamp.
4
Auto irradiation (-)
-
5
Lamp ON signal
OUT
Outputs a signal (open collector output) as long as
the lamp is lit.
6 Error
signal
OUT
Outputs a signal (open collector output) when an
error such as a lamp error, lamp lifetime error and
fan error occurs.
7
N.C.
-
Do not connect.
8
N.C.
-
Do not connect.
9 +24
OUT
+24 V DC output (100 mA max.)
10 +24 OUT
11 +24 OUT
12 GND -
Ground for external control connection
13 GND -
14 GND -
15 GND -
Connector external view and pinout
Содержание EX-mini L12530
Страница 21: ...21 11 Dimensional outlines Unit mm...