Page 19 of 33
V1.1
2.3.5 Modifying Parameters (Settings)
Particle Sizes:
This six-channel particle counter allows the user to select six channel sizes from 0.3
μm – 5.0 μm in steps of 0.1 μm and from 5.0 μm – 25 μm in steps of 1 μm. Channel
1 will always be the smallest size, followed by channel 2, 3, and so on. If the
FS209E/ISO alarm is enabled, the lowest two channels will be locked to 0.3 μm and
0.5 μm and cannot be changed until the FS209E/ISO alarm has been disabled.
Sample Parameters:
The sample parameters screen will allow the user to modify the following parameters
over the following ranges:
•
Start Delay
:
00:00
to
59:59
•
Sample Interval
:
00:00
to
59:59
•
Sample Duration
:
00:00
to
59:59
•
Number of Samples
:
0
to
500
•
Material
:
PSL
,
Sand
,
Lead
,
Carbon
, or
User Defined
•
Count Mode
:
Cumulative
,
Differential
,
Counts per Liter
,
Counts per Cubic
Foot
,
Counts per Cubic Meter
,
Mass Concentration
(option) or
Particulate
Matter
(option)
•
Location
:
A0
to
Z9
Start Delay
The start delay is defined as the time before ANY measurements are taken. This
feature is a timer for a sampling process to occur at a later time. The sample
Particle Sizes
Sample Parameters
Date & Time
Preferences
Alarm/Sound Levels
Print/Save Options
Bluetooth Settings
Factory Settings