TESTit
Faults
Order h49 7144.907-333
115
12.2 Fault table
Fault
Possible cause
Remedial action
Remedied by
IT module is not
found, the gray
Bluetooth symbol
is not displayed
IT module's re-
chargeable battery
is discharged
Charge IT module
Skilled special-
ist
Bluetooth driver of
the display device
is not working cor-
rectly
Restart the display device and the
TESTit software
Skilled special-
ist
Bluetooth is deac-
tivated on the dis-
play device
Activate Bluetooth in Windows
settings
Skilled special-
ist
IT module not in
range
Move IT module closer to TESTit
software [display device]
Skilled special-
ist
Unable to connect
to IT module alt-
hough it is de-
tected and listed
Rechargeable bat-
tery of the IT mod-
ule is too weak
Charge IT module
Skilled special-
ist
IT module not at-
tached to TEST
module
Attach IT module to TEST module
Skilled special-
ist
IT module is at
range limit of the
radio signal
Move IT module closer to TESTit
software [display device]
Skilled special-
ist
IT module needs to
be rebooted
Reboot the IT module [see chap-
ter »Resetting the IT module«]
Skilled special-
ist
Gray Bluetooth
symbol is
displayed, but IT
module is not
shown in the
selection window
IT module was not
visible for Blue-
tooth driver at the
time the gray Blue-
tooth symbol was
pressed
Close and reopen the »Select
clamping force measuring device«
window by pressing »Cancel«.
Skilled special-
ist
Connection to the
TESTit V1 device
cannot be
established
TESTit V1 device
is out of Bluetooth
range
Bring the TESTit V1 device within
range of the software and connect
again
Skilled special-
ist
Display device
does not start
Display device's
rechargeable bat-
tery is discharged
Charge the rechargeable battery
of the display device
Skilled special-
ist
Display device has
a defect
Return the display device to the
manufacturer
HAINBUCH