64
Software
of the ADC, where 2
23
(8388608) is the maximum of modulation. The X
– Axis shows the number
of recorded samples. The Sampling rate is 48 kHz, it corresponds with the time, where the unit 1
is 20.83 μs.
Button description:
Up / Down Amplitude
With Up, the signal amplitude is shown larger. The Auto Scale button will
be automatically disabled.
With Down, the signal amplitude is shown smaller, this increases the scale.
Auto Scale
Setting this option will show the signal well scaled.
The voltage over the nominal standard capacitor will be shown
The voltage over the test object will be shown
The mains input voltage will be shown
Inc / Dec Time Base
By pressing "ms" Button, the time scale will be increased. As a result, there
are more recorded data visible.
By Pressing "
s" Button, the time scale will be decreased. As a result,
there are less data will be shown.