Device Functions
21
I
X
Current trough Device Under Test C
X
I
N
Current
trough
known
Standard
Capacitor C
N
I
RX
Losses of the Device Under Test C
X
C
X
Test Object (ideal capacitance)
C
N
Standard capacitor (with tan
< 10
-5
)
R
X
Measuring shunt for I
X
, C
X
R
N
Measuring shunt for I
N
, C
N
V
Low voltage point of the HV supply and
reference point of the measurement
ADC Analogue to Digital Converter
t
1
, t
2
Time stamps of the measured values
6.1.2
DC Resistivity
The DC Resistivity measurement is based on a pico ampere meter built-in the 2831 which
measures the current through the liquid or solid test cell. The HV DC source built-in the 2831
energizes the test cell so that a current can flow through the solid or liquid sample and the pico
ampere meter. The applied voltage and the measured current are converted to digital values for
further calculations in the 2830.
To calculate the resistivity, four parameters has to be known: The Ix through the solid or liquid
sample, the applied voltage Utest, the surface area of the measuring electrode A and the thickness
of the test sample l. In the chapter 9.5 DC Resistivity the calculation is explained in detail.
Содержание 2830
Страница 1: ...2830 2831 Precision Liquid and Solid Dielectric Analyzer User Manual 4843477 M Version 3 0 ...
Страница 2: ...II Revision History Version Date Author Remarks 2 0 2013 12 MB 2nd release 3 0 2021 07 SM 3rd release ...
Страница 7: ...Introduction VII 17 Conformity 113 18 Notes 114 ...
Страница 8: ...VIII Introduction ...
Страница 121: ...Conformity 113 17 Conformity ...
Страница 122: ...114 Notes 18 Notes ...
Страница 123: ...Notes 115 ...