60
Software
The measured values are not yet stable. The built-in averaging routine is still
calculating a mean value. Normally during voltage or frequency changes the
values are displayed in dark green.
The value is now stable and has the correct accuracy. In an automatic sequence
mode the value will be recorded now and the next voltage step will be set.
Signal Overflow error. This case should not be seen during normal operation.
Except the first element (U rms) the other three measurement displays can be customized by
clicking the arrow
in the top right corner of the element.
Following measuring values can be selected:
DF (tan
)
Actual measured dissipation factor
DF%(tan
)
Actual measured dissipation factor in percentage format
Electrical Stress
Actual measured electrical stress. The electrical stress is
depending on the voltage and the gap between the electrodes.
The gap for the liquid test cell is hard implemented to 2 mm and
for the solid test cell the gap is given in the text field “Plate
Distance” in the sub tab sheet “Settings” (see chapter 9.3.2
Settings)
Permittivity
Actual measured relative permittivity
U rms
Effective voltage applied to the test object
Frequency
Measured frequency of applied voltage
I
n rms
Effective current through C
N
I
x rms
Effective current through C
X
Cx
Value of measured capacitance
Содержание 2830
Страница 1: ...2830 2831 Precision Liquid and Solid Dielectric Analyzer User Manual 4843477 M Version 3 0 ...
Страница 2: ...II Revision History Version Date Author Remarks 2 0 2013 12 MB 2nd release 3 0 2021 07 SM 3rd release ...
Страница 7: ...Introduction VII 17 Conformity 113 18 Notes 114 ...
Страница 8: ...VIII Introduction ...
Страница 121: ...Conformity 113 17 Conformity ...
Страница 122: ...114 Notes 18 Notes ...
Страница 123: ...Notes 115 ...