62
Software
Set Voltage
In this text field the AC test voltage can be set.
If the option “Touch screen” is set in the sub tab sheet
Options (see 9.3.4 Options) a voltage input dialog will be
opened. The info text on the right side will be displayed
where the range of the settable voltage is specified.
If the option “external” is selected then no voltage input dialog
will be opened. The voltage can be typed directly into the text
field.
If a voltage higher than 2 kV is set then a warning message
will pop up because the maximal AC voltage for the liquid
and solid Tettex test cells is 2 kV.
If a voltage out of range is set this info message will pop up.
Record
The actual selection of measured data will be stored in the
spreadsheet.
Only the following three states of the HV ON/OFF button are described here. The other buttons
are descried in the chapter 9.1.4 Basic Buttons.
HV Disabled
When the HV ON/OFF Button is disabled then no heater or test cell is
selected, the emergency button is pressed, one of the interlock
connectors are not connected or another measurement is running.
HV ON
This button state is used to switch ON the set voltage.
HV OFF
This button state is used to switch OFF the set voltage.
Содержание 2830
Страница 1: ...2830 2831 Precision Liquid and Solid Dielectric Analyzer User Manual 4843477 M Version 3 0 ...
Страница 2: ...II Revision History Version Date Author Remarks 2 0 2013 12 MB 2nd release 3 0 2021 07 SM 3rd release ...
Страница 7: ...Introduction VII 17 Conformity 113 18 Notes 114 ...
Страница 8: ...VIII Introduction ...
Страница 121: ...Conformity 113 17 Conformity ...
Страница 122: ...114 Notes 18 Notes ...
Страница 123: ...Notes 115 ...