AXOS
5
Compact immunity test System
1
1
Safety
1.1
General Safety Information
Significant: Please read carefully the safety requirements before using the AXOS
5
!
This warning sign is visible on the equipment.
Meaning: The documentation must be consulted before connecting the equipment to any voltage
supply
Dangerous mains voltage or high voltages are present inside the AXOS
5
.
The earth connection of the AXOS
5
must be connected to a good earth.
If trigger is pressed without connection to earth, electrical shock may occur!
Before removing any covers, remove all external connection cables.
The AXOS
5
should only be maintained by trained personnel.
Do not open any system modules, they contains no user replaceable parts.
People with heart pacemakers must not be in the vicinity of the system, when it is in operation.
Do not switch on or operate the AXOS
5
System if an explosion hazard exists. The system should be
operated in a dry room. If condensation is visible the unit should be dried before operating.
Never touch the Equipment Under Test (EUT), when the AXOS
5
is operating. Establish a safety barrier
around the test setup (Close the circuit pin 4 and pin 5 in Table 5-3).
If any part of the AXOS
5
is damaged or it is possible that damage has occurred, for example during
transportation, do not switch on the unit.
Before opening any units remove the mains power cord.
Before changing the mains fuse, remove the mains power cord.
Fuses should only be replaced with the same type and value.
This user manual is an integral part of the test system. Haefely Test AG and its sales partners refuse
to accept any responsibility for consequential or direct damage to persons and/or goods due to none
observance of instructions contained herein or due to incorrect use of the AXOS
5
.
When applying Magnetic field, do not touch the coil when test is running. Only direct output HI and
COM must be used to connect the coil. The coil loop (from HI till COM) must be closed.
Never connect a supply voltage directly to the HI/COM surge outputs. This will damage the unit!!!.
Always use a suitable CDN
Table 1-1: Safety requirements AXOS
When a test is started on the AXOS
5
the symbol in Figure 1-1 is
displayed on the screen.
Figure 1-1: Safety sign
1.2
Safety Standards
The AXOS
5
fulfils the requirements of IEC 61010-1.
Содержание AXOS5
Страница 1: ......
Страница 6: ......