![Haefely Technology AXOS5 Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/haefely-technology/axos5/axos5_user-manual_577511014.webp)
8
AXOS
5
Compact immunity test System
5
General operation
5.1
Front view
The Figure 5-1 shows the front view of the AXOS
5
. In the Table 5-1 is the detail description of every physical input and
output.
1
2
3 4
6
13
7
8
9
11
14
5
12
10
Figure 5-1: AXOS
5
Front view
Table 5-1: Description front view function
Pos.
Function
Description
1
On/off switch
Turns on and off the power of AXOS
5
2
Surge impulse voltage
monitor output
1000:1 divider for surge impulse.
Direct connection to oscilloscope.
Example: 5000Vsurge impulse output will show 5V on this
BNC socket
3
Surge impulse current
monitor output
1000A/V surge impulse current monitor.
Direct connection to oscilloscope.
Example: 2500A surge current will show 2.5V on this BNC
socket
4
Urms monitor output 100:1
5
Irms monitor output 10A/V
Monitoring the signal of the generator when creating voltage
dips or voltage interrupts.
Direct connection to oscilloscope.
6
HI Generator output Surge; (Direct
output)
7
COM Generator output Surge
(Direct output)
Direct output of surge generator
Connection to external CDN.
Do not connect directly any external supply voltage, generator
will be damaged.
8
Burst Output
Direct output of burst generator to external CDN or EUT
9
L output to EUT
10
N output to EUT
11
PE output to EUT
Connection for single phase EUT to the internal
coupling/decoupling network
12
PE
Connection to ground reference plane or the verification
adater
13
USB port for data transfer
Save report data or test files onto USB memory stick
14
Colour touch screen
Every command and adjustment will be controlled and entered
via the touch screen
Содержание AXOS5
Страница 1: ......
Страница 6: ......
Страница 49: ...AXOS 5 Compact immunity test System 43 15 Appendix 15 1 CE conformity ...