AXOS
5
Compact immunity test System
23
The following graphic represents the pre-compliance mode. In comparison with the standard menu, every command
gets displayed the entire visual description of the function is described through purple bows. Main advantage of the
pre-compliance mode is that the parameter can be changed while the test is operating.
Figure 8-3: EFT / BURST pre-compliance mode
However, when every parameter has been adjusted, the “Start” button can be pushed and generator is charging. Now,
the following graphic should be displayed. The selected output path (direct or line) is flashing yellow.
Figure 8-4: Burst pre-compliance mode charging
The Burst waveform gets stopped, when pushing the “Stop” button. Afterwards, test report can be saved on the USB
device.
8.2.1
Output & Coupling Paths
In Burst testing mode every line signal gets tested against GND (PE). In direct output the signals stands on the
position 8 in Figure 5-1. It gets connected directly to the EUT or will be used for verification of the Burst waveform.
Furthermore, it can be connected to three-phase test CDN. For further assistance please contact the HAEFELY TEST
AG directly.