AXOS
5
Compact immunity test System
11
Menu field
Description
Language
Selection of different language, in the moment only English available
Date & Time
Entering of current date and time
Sound Volume
Adjusting of sound operating volume and warn signals
Report data
saving
Reporting data can be automatically saved when pushing the device in USB port or manually by
pressing the “Rep.Data” in the menu of every application and then entering the device
Start file
Selection of predefined test in accordance with particular norm for Surge, Burst, Magnetic field
and Voltage dips & interrupts; it gets only applied from an external start/stop automatic (
Last Calibration
Factory calibration from HAEFEY TEST AG, in future term the “next calibration” becomes “last
calibration
Next calibration
Recommended calibration period from HAEFELY TEST AG is 2 years, however, it always
depends on the quality management system of every company and certainly the test
requirements of equipment and certification environment
Remote GUI
Remote GUI is possible to access when license key purchased from the HAEFELY TEST AG;
switch on/off of communication between remote computer and AXOS5 possible
Coil Antenna
Entering of new “coil Antenna” factor as described in
Communication
Communication via RJ45 at the rear view of AXOS
5
Firmware update
For updating the new operating software of the AXOS
5
License Manager
Entry of license codes for Surge, Burst, Magnetic field and Voltage Dips & Interrupts and remote
GUI. After entering the license code, accessible in main menu
Touch calibration
In normal delivery mode there is a pre-calibration of the touch screen. However, if required,
please follow and press for few seconds with a pen the cross at the screen. It requires five
different positions. Then confirming and as a control measure four different yellow dots get open.
Again confirming and the calibration has finished.
Table 5-4: Setup description
After setting has been finished the “Setup” gets closed when pressing the “close” button. Afterwards the starting
window gets displayed. It can be selected between Surge, Burst, Magnetic field and Voltage Dips & Interrupts.
5.4
Predefined standard program
The AXOS
5
contains predefined standards test for “Surge”, “Burst”, ”Magnetic field” and “Voltage Dips & Interrupts”
according to each particular norm. Each test can be selected and applied through the Generator to the EUT. The
Figure 5-4 reflects the save and option of data at the “Surge” generator. These application works similar for “Burst”,
“Magnetic field” and “Voltage Dips & Interrupts. The basic understanding of the tool is similar to normal office
applications. Data can be “load”, “Save”, “Save as”, “Copy” and “Delete” as for instance in word. These options occur
on the screen, when “file” button is pushed.
Figure 5-4: Save and load option of data
Two directories can be selected as shown in Figure 5-5. However, the “standard” directory can only read and load the
predefined test procedure in accordance to the particular norm. These files can only be read. The second directory is
“Internal”. The internal data (customized testing files) can be saved, loaded and deleted. Either of the directories can
only be accessed out of the particular menu, like for instance the “surge” generator.
In addition, the loaded program is displayed in the upper line in either of the standard and pre-compliance menu.