AXOS
5
Compact immunity test System
21
7.1.2
Transition
In “Transition” (Figure 7-5) the user can adjust the “Field Strength”, “Sync. Phase” and “Alternative Polarity”. The
description from capture 6.3.3 is valid, except “peak voltage” has been changed to “field strength”. The other
parameters stay equal, since an OCC will be applied to the coil antenna and is basically a Surge impulse.
Figure 7-5: Magnetic field transition
When all parameters has been entered then simply confirming with ok and in the menu is showing “transition on”, this
indicates the settings will be applied to the coil antenna when pushing the “start” button. Furthermore, when test is
proceeding, the intervals for instance of the “field strength” gets displayed.
7.1.3
Properties
Position
description
External (Pin11 on Aux)
External condition (e.g. device or PLC signal), the
position becomes 1 and causes an action (Ignore,
Alarm, Test Stop or test stop & line off)
Line Current
Limits the current (L, N, PE front view) to the EUT.
Action: Ignore, Alarm, Test stop, Test stop & line off
Due to EUT fails, AXOS
5
causes an action which can
be: Ignore, Alarm, Test Stop, Test Stop & Line off.
Line off
Voltage locked at L, N, PE when test has finished
Line on
Voltage unlocked at L, N, PE when test has finished
Sound on or off
Acoustic signal when test ends
Table 7-1: Magnetic field properties