AXOS
5
Compact immunity test System
25
8.2.4
Trigger
When selecting “manual trigger” the following window opens up. It gets applied to EUT by pressing the “Trigger”
button. It gets always only one Burst package applied.
Figure 8-7: BURST mode Trigger normal
Furthermore, it shows between which line and GND the burst package gets applied. After finalizing the process, it will
be stopped when pushing the “Stop” button. A manual report gets saved on the USB device, after selecting “Rep.Data”
and “manual” selection in the “Setup” menu.
8.2.5
Synchronization
To synchronize the burst signal with the main supply source it is necessary to select in the menu “Synchronization” and
then “Line Sync.”. Adjustments of the angle can be done in a range from 0° to 359°. However, if no power supply is
connected to the EUT supply input at the rear view, it has to be entered “Async.”, otherwise it is impossible to get a
Burst waveform.
In external synchronization, the output with the position 8 Figure 5-2 in the rear view is in use. An external source could
be for instance a manual CDN. Finally, normal operation through the touch screen is possible.
8.2.6
Burst mode
The Burst mode gives the user the option to select between “Normal”, “Continuous”, “Real” and “Random BURST”
signals. Afterwards the function gets automatically displayed on the screen and can be applied to the EUT after
pushing the “Start” button.