AXOS
5
Compact immunity test System
The safety symbol, the direct output and the output voltage and output current are displayed. In addition, the user is
able to see nominal values of the peak voltage, reputation time and the synchronization period. Additionally, these
values can be modified while the test is proceeding. It happens by simply pressing each particular function. The test
stops when pushing the “STOP” button.
Figure 6-6: Surge Pre-compliance mode charging
In the right corner of the graphic is the plug flashing in red, when direct connection between L, N, PE to the “EUT
Supply Input” and button “line” has been pushed. The safety lock symbols shows that the safety circuit has been
closed between PIN 4 and PIN 5 in Table 5-3.
Information about Transition, Trigger and properties has to be entered independently if the test has stopped.
“Transition” and “Trigger” are only accessible in standard operating mode.
6.3.1
Output & Coupling Paths
In the IEC norm 61000-4-5 it is written that every coupling phase has to be tested against each other. That realizes the
AXOS
5
with the line output and selecting the options independently. The HI and Com output gets active when in
operating menu “direct” output is selected. L, N, PE can be selected when in “line”.
6.3.2
Properties
In the Figure 6-7 below is the properties menu displayed. The Table 6-2 explains the function more detailed.
Figure 6-7: Surge properties
The Table 6-2 describes the different functions more detailed. The window gets closed, when confirming the “OK”
button.