6
AXOS
5
Compact immunity test System
4
Initial operation
4.1
Visual check
During transport AXOS
5
may be subjected to excessive shocks and vibrations, even though every care is taken by
HAEFELY Test AG to provide suitable packaging. Before operating the unit, check for signs of mechanical damage.
Damaged packing cases may by a sign of transport damage. Damage caused in transit must be reported to the
shipping agent immediately.
4.2
Installation
All safety measures must be followed, as described in Table 1-1.
4.3
Earth connection
The "Earth Stud" on the back panel must always be connected to a solid earth before the "EUT Supply
Input" is connected to power source. Cross section of connection must be at least 2.5mm squared.
4.4
EUT Supply Input
The "EUT Supply Inpu" on the back is the connection to the integrated single-phase CDN, which can be used
up to 16A / 264V.
There are three banana-plugs (stackable) delivered with the standard accessories set:
- black: line conductor
- blue: neutral conductor
- green/yellow: protection earth conductor
These banana plugs shall be mounted on a suitable connection cable.
The EUT Supply Input must be connected to power lines which are protected by a line safety switch
equal or less 16A rated current and characteristic B or C.