Calibration
73
(
8.1.2.1
Calibration in air (recommended)
1.
Press
CAL
, enter passcode:
1100
,
ENTER
. The display will briefly
show CAL and then the entered (or default) humidity value.
2.
Remove the sensor from the process and place in air.
3.
Enter the actual relative humidity and press
ENTER
. The display will
show the entered (or default) barometric pressure.
4.
Enter the barometric pressure and press
ENTER
. A drift check will
begin. The drift check may take some time. When complete, the
display will show the concentration in the units specified in the
configuration menu.
Note: To exit the drift check after, 10 seconds, press CAL (accuracy
will be reduced).
5.
Accept the calculated value for the concentration or adjust the value
as necessary and then press
ENTER
. The display will show the new
slope and zero (relative to 25 °C and 1.013 mbar).
6.
Press
ENTER
to end the calibration. Return the sensor to the
process. The measured value will be shown in the display
alternately with Hold.
7.
Press
ENTER
to return to the measure mode.
Содержание si792 D series
Страница 2: ...2 ...
Страница 6: ...Table of Contents 4 ...
Страница 12: ...Specifications 10 ...
Страница 36: ...Interface and navigation 34 ...
Страница 44: ...Operation si792 x D 4 20 mA HART 42 ...
Страница 66: ...Operation Foundation Fieldbus 64 ...
Страница 72: ...Operation Profibus PA 70 ...
Страница 80: ...Maintenance 78 ...
Страница 86: ...Troubleshooting 84 ...
Страница 88: ...Parts and accessories 86 ...
Страница 93: ...Index 91 Z zero adjustment 72 ...
Страница 94: ...Index 92 ...
Страница 95: ...93 ...