Operation—si792(x) D (4–20 mA/HART)
36
To exit the configuration mode at any time, press
CONF
and then
ENTER
. The output current will be held for 20 seconds and the
measured value will be displayed.
Note: During configuration the transmitter remains in the Hold mode for
safety reasons. The loop current is frozen at the value specified in the
o1.HoLD menu option. The Sensoface icon is inactive. The
configuration mode indicator is displayed (
).
5.2.1
Output configuration
o1
4–20 mA current output
1200
tc
Temperature sensor
Corr
Correction for pressure,
salinity, polarization voltage.
CAL
Calibration setup
AL
Alarm settings
Table 9 Output setup menu—out.1
MNU
Select the parameter
o1. FCT
DO (concentration by weight; default)
GAS (concentration by volume)
Select the sensor type
o1.SnSR
Type A (standard; default) examples: Oxysens, Oxyferm
Type B (trace measurements) example: Oxygold G
Select the units
o1.UniT
DO (default)
SAt % (percent saturation; default)
Conc µg/L
Conc mg/L
Conc ppb
Conc ppm
GAS
GAS % (default)
GAS ppm
Table 8 Configuration menu
Содержание si792 D series
Страница 2: ...2 ...
Страница 6: ...Table of Contents 4 ...
Страница 12: ...Specifications 10 ...
Страница 36: ...Interface and navigation 34 ...
Страница 44: ...Operation si792 x D 4 20 mA HART 42 ...
Страница 66: ...Operation Foundation Fieldbus 64 ...
Страница 72: ...Operation Profibus PA 70 ...
Страница 80: ...Maintenance 78 ...
Страница 86: ...Troubleshooting 84 ...
Страница 88: ...Parts and accessories 86 ...
Страница 93: ...Index 91 Z zero adjustment 72 ...
Страница 94: ...Index 92 ...
Страница 95: ...93 ...