Operation—Foundation Fieldbus
54
6.2.2.6 Cyclic measured value status
The cyclic measured value status is shown in
.
6.2.2.7 Measured value limits—limit bits
The respective status bit is set when a condition occurs (
). The
status bit is reset when the condition no longer exists.
Table 17 Cyclic measured value status
Priority
Quality
Sub-status
Bin-coding
(no limit bits)
Hex-
coding
Low
Good
Good Non-Specific
10 00 00 00
0 x 80
Good Active Advisory Alarm
10 00 10 xx
0 x 88
Good Active Critical Alarm
10 00 11 xx
0 x 8C
Uncertain
Uncertain Non-Specific
01 00 00 xx
0 x 40
Last Usable Value (LUV)
01 00 01 xx
0 x 44
Substitute-Set
01 00 10 xx
0 x 48
Initial Value
01 00 11 xx
0 x 4C
Sensor Conversion Not Accurate 01 01 00 xx
0 x 50
Engineering Unit Violation
01 01 01 xx
0 x 54
Sub-Normal
01 01 10 xx
0 x 58
Bad
Non-Specific
00 00 00 xx
0 x 00
Sensor Failure
00 01 00 xx
0 x 10
Device Value
00 00 11 xx
0 x 0C
High
Out of Service
00 01 11 xx
0 x 1C
Table 18 Limit bit description
Bin coding of limit bits
Description
00
OK
01
Low-limited
10
High-limited
11
Constant
Содержание si792 D series
Страница 2: ...2 ...
Страница 6: ...Table of Contents 4 ...
Страница 12: ...Specifications 10 ...
Страница 36: ...Interface and navigation 34 ...
Страница 44: ...Operation si792 x D 4 20 mA HART 42 ...
Страница 66: ...Operation Foundation Fieldbus 64 ...
Страница 72: ...Operation Profibus PA 70 ...
Страница 80: ...Maintenance 78 ...
Страница 86: ...Troubleshooting 84 ...
Страница 88: ...Parts and accessories 86 ...
Страница 93: ...Index 91 Z zero adjustment 72 ...
Страница 94: ...Index 92 ...
Страница 95: ...93 ...