Operation—Foundation Fieldbus
58
APPLICATION
Selects the measurement
method
(DO/ GAS)
R/W 2
uns16
0 = Dissolved
oxygen
1 = Measurement in
gases
PRIMARY_VALUE
Shows the primary value and
status
Value
Status
R
4
1
DS_65
PRIMARY_VALUE_TYPE
Selects the displayed primary
value depending
on the selected measuring
method in APPLICATION
parameter
R/W 2
uns16
0 = DO_%
1 = DO_mg/l
2 = DO_
g/l
3 = DO_ppm
4 = DO_ppb
5 = GAS_ppm
6 = GAS_%
FILTER_TIME
Sets the filter time of the input
filter
R/W 2
uns16
0–120
Default: 0s
IMPEDANCE
Shows the impedance of the
sensor
R
4
float
SENSOR_CURRENT
Shows the momentary current
of the sensor
R
4
float
Temperature
Table 19 Transducer block bus parameters (continued)
Parameter
Description
R/W Bytes
Data
type
Range