Operation—Profibus P
A
68
Memory error
(RAM, ROM,
EPROM)
3
0000 11xx
Failure
ERR MEMORY
Yes
Measurement range
violation
4
0100 0111
0100 1111
Failure
ERR SAT VALUE
Yes
Measurement range
violation
5
0100 0111
0100 1111
Failure
ERR CONC VALUE
Yes
Temp range violation
Temperature sensor
6
0100 0111
0100 1111
Failure
ERR TEMP VALUE
Yes
Membrane defective
7
0100 0111
0100 1111
Failure
ERR SENSOCHECK
Yes
Zero point/Slope
8
0101 00xx
Maintenance
req.
CHK ZERO/SLOPE
Yes
Sensor response time 9
0101 00xx
Maintenance
req.
CHK EL. RESPONSE
Yes
Calibration timer
Cal prompt
10
0101 00xx
Maintenance
req.
CAL REQUIRED
Yes
Calibration
9
0100 0111
0100 1111
Function check CAL RUNNING
Yes
Configuration
10
1010 00xx
Function check CONF RUNNING
Yes
Table 20 PROFIBUS PA communication (continued)
Cause
No. of
binary
message
Analog input
status
Physical Block
(PB)
Global status
Text of binary message
(default)
Logbook
(default)
Содержание si792 D series
Страница 2: ...2 ...
Страница 6: ...Table of Contents 4 ...
Страница 12: ...Specifications 10 ...
Страница 36: ...Interface and navigation 34 ...
Страница 44: ...Operation si792 x D 4 20 mA HART 42 ...
Страница 66: ...Operation Foundation Fieldbus 64 ...
Страница 72: ...Operation Profibus PA 70 ...
Страница 80: ...Maintenance 78 ...
Страница 86: ...Troubleshooting 84 ...
Страница 88: ...Parts and accessories 86 ...
Страница 93: ...Index 91 Z zero adjustment 72 ...
Страница 94: ...Index 92 ...
Страница 95: ...93 ...