Specifications
8
Profibus-PA (DPV1)
Bus-powered device with constant
current consumption. Cyclic and acyclic
data exchange. Physical block,
2 analog input function blocks,
2 discrete input blocks, logbook block,
alarm block.
PNO directive: PROFIBUS-PA,
Profile for Process Control Devices,
Version 3.0
Physical interface: MBP-IS
(Manchester Bus Powered-Intrinsically
Safe) to IEC 1158-2 (DIN-EN 61158-2)
Connection: via segment coupler to
SPC, PC, PCS
Address range: 1 to 126
Temperature Input
Probe
NTC 30 k
/NTC 22 k
(selectable),
2-wire connection
Range
–20.0 to 150.0 °C (–4 to 302 °F)
Adjustment range
10 K
Resolution
0.1 °C; 0.1 °F
Measurement error
1
,
2
<0.5 K
(<1 K at T >100 °C)
Dissolved oxygen input
Type A sensor
3
Oxyferm, Oxysens (Hamilton)
Type B sensor
3
Oxygold G (Hamilton)
Measuring current
0–1200 nA; resolution: 20 pA
Measurement error
1
,
2
0.5% measured value + 0.05 nA
Temperature coefficient
0.005 nA/K
Permitted guard current
20 µA
Polarization voltage
400–1000 mV; resolution:
approximately 3 mV
Range, percent saturation
0–199.9%/200–500% (–10 to 80 °C)
Содержание si792 D series
Страница 2: ...2 ...
Страница 6: ...Table of Contents 4 ...
Страница 12: ...Specifications 10 ...
Страница 36: ...Interface and navigation 34 ...
Страница 44: ...Operation si792 x D 4 20 mA HART 42 ...
Страница 66: ...Operation Foundation Fieldbus 64 ...
Страница 72: ...Operation Profibus PA 70 ...
Страница 80: ...Maintenance 78 ...
Страница 86: ...Troubleshooting 84 ...
Страница 88: ...Parts and accessories 86 ...
Страница 93: ...Index 91 Z zero adjustment 72 ...
Страница 94: ...Index 92 ...
Страница 95: ...93 ...