Operation—Foundation Fieldbus
45
Co.PrES
0.000–9.999 BAR (default: 1.013 BAR)
999.9 kPa / 145.0 PSI (not in sensor)
Select salinity correction
Co.SAL
0.00–45.00 mg/L (default: 0.00 mg/L)
Select calibration mode
Ca.MOD
SAt (default)
Conc
Select calibration interval
CA.tiME
0000–9999 h (default 0000 h)
Select Sensocheck
AL.SnSO
CHECK ON
Continuous Sensocheck evaluation of
sensor function
CHECK OFF (default)
LED in Hold mode
AL.LED
HOLD ON
LED blinks during hold
HOLD OFF (default)
LED off during hold
Select Fieldbus address (optional)
1
FF.ADR
0017–0031 BUS (default: 0026 BUS)
1
Use only when there is no bus connection. The transmitter will restart and set all
parameters to default values. Individual settings must be entered once more.
6.1.2
Configuration menu (continued)