Normal operation of this device may require the use of chemicals or samples that are biologically
unsafe.
• Observe all cautionary information printed on the original solution containers and safety data
sheets prior to their use.
• Dispose of all consumed solutions in accordance with the local and national regulations and laws.
• Select the type of protective equipment suitable to the concentration and quantity of the dangerous
material being used.
If the wiper interval is correctly set for the application and the wiper profile is replaced regularly, more
cleaning of the measuring path is not necessary.
1.
Go to the main menu, then select
Sensor setup
.
2.
Select the applicable sensor, if necessary.
3.
Select
Maintenance
.
4.
Set the outputs on
Hold
.
5.
Remove the probe from the tank or flow through unit.
6.
Go back to the sensor menu, then select
Diagnostics/Test
>
Signals
.
7.
Rinse the probe with clean water.
As an alternative, use a window cleaner, grease remover or 5 % hydrochloric acid to clean the
measuring path based on the quantity and type of contamination. Soak the probe in the cleaning
solution for 5 to 10 minutes.
8.
Carefully clean the measuring path with distilled water.
Clean the measuring path until the
DExt1
and
DExt2
values are less than 15.
9.
Select
Back
to go back to
Diagnostics/Test
.
10.
Select
Back
again.
11.
Select
Return sensor to process
.
The probe goes back to measurement mode after the automatic wipe cleaning.
12.
Install the probe in the tank or flow through unit.
6.3 Replace the wiper blade
Replace the wiper blade when one or more of the conditions that follow occurs:
• After 25,000 cleaning cycles (A-B-A or B-A-B)
• After 50,000 cleaning single movements (A-B or B-A)
• if the wiper blade has damage or does not operate correctly.
1.
Remove the probe from the tank or flow through unit.
2.
Go to the maintenance menu as follows:
• SC200 and SC1000 Controllers—Go to the main menu, then select
Sensor setup
>
Maintenance
.
• SC4500 Controller and Claros interface—Push the tile of the device and select
Device menu
>
Maintenance
.
3.
Select
Wiper replacement
.
4.
Complete the instructions on the display. When asked, replace the wiper blade. Refer to the
illustrated steps in
and
.
English
19
Содержание NT3100sc
Страница 2: ...Table of Contents English 3 Fran ais 27 Espa ol 52 Portugu s 78 102 124 147 169 2...
Страница 11: ...Figure 3 Bracket installation English 11...
Страница 12: ...Figure 4 Flow direction top view Figure 5 Sensor with flow through unit installation 12 English...
Страница 20: ...Figure 8 Replace wiper blade 1 mm and 2 mm Figure 9 Replace wiper blade 5 mm 20 English...
Страница 36: ...Figure 4 Sens du d bit vue du dessus Figure 5 Installation d un capteur avec unit de d bit 36 Fran ais...
Страница 61: ...Figura 4 Direcci n del flujo vista superior Figura 5 Instalaci n del sensor con celda de flujo Espa ol 61...
Страница 86: ...Figura 3 Instala o do suporte 86 Portugu s...
Страница 87: ...Figura 4 Dire o do fluxo vista superior Figura 5 Instala o do sensor com a unidade de fluxo direto Portugu s 87...
Страница 104: ...2 1 2 1 1 2 1 2 104...
Страница 108: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5 mm 5 2 9 1 mm 2 mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 4 9 108...
Страница 109: ...3 109...
Страница 110: ...4 5 110...
Страница 118: ...9 5 mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 118...
Страница 126: ...2 1 1 2 1 2 126...
Страница 130: ...2 6 2 2 1 NT3100sc NT3200sc 3 5 mm 5 9 2 9 1 mm 2 mm 5 4 LZY261 3 3 1 9 130...
Страница 131: ...3 2 3 5 4 3 131...
Страница 132: ...4 5 132...
Страница 133: ...3 3 SC 6 6 4 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 133...
Страница 135: ...Off Auto Selection 3 6 12 24 1 3 1 2 3 4 5 3 5 3 1 1 SC200 SC1000 SC4500 Claros 2 3 Offset 10 10 0 124 0 8 1 2 1 135...
Страница 136: ...Start calibration Standard calibration 1 4 3 6 5 3 2 NT3200sc 1 2 3 15 1 2 3 3 NO3 NO2 NO3 1 3 146 1 1 2 3 3 1 2 ID 136...
Страница 140: ...8 1 mm 2 mm 9 5 mm 140...
Страница 142: ...10 7 5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 UNSTABLE 142...
Страница 143: ...5 6 NO3 NO3 N NO3 NO3 N NO3 NO3 N NO2 NO2 N NO2 NO2 N NO2 NO2 N 8 Web 143...
Страница 149: ...2 1 1 2 1 2 UV 149...
Страница 151: ...2 3 NT3100sc NT3200sc 2 4 250nm 2 5 NT3100sc NT3200sc 1 TSS3 SC SC 1 2 1 NT3100sc 1mm 2mm 5mm 5 000mg TSS L 3 TSS 151...
Страница 153: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5mm 5 2 9 1mm 2mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 9 153...
Страница 154: ...4 3 154...
Страница 155: ...4 5 155...
Страница 156: ...3 3 SC 6 6 4 ON 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 156...
Страница 158: ...3 6 12 24 1 3 1 2 3 4 5 3 5 3 1 1 SC200 SC1000 SC4500 Claros 2 3 10 10 0 147 0 8 1 2 1 1 1 4 3 6 158...
Страница 163: ...9 5mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 163...
Страница 164: ...c 10mL d 8 SC200 SC1000 SC4500 Claros Enter 9 NT3200sc NT3200sc NT3200sc 10 11 12 10 7 5 164...
Страница 165: ...5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 6 165...
Страница 171: ...5 1 2 PUR HT CE CMIM UKCA FCC ISED 1 2 2 2 1 2 1 1 171...
Страница 172: ...2 1 2 UV 2 1 3 2 1 4 ICES 003 Class A Class A 172...
Страница 175: ...1 1 3 2 2 6 2 2 1 NT3100sc NT3200sc 3 9 5 5 2 9 1 2 5 4 LZY261 9 175...
Страница 176: ...3 3 1 OFF 3 2 3 5 4 176...
Страница 177: ...3 177...
Страница 178: ...4 5 178...
Страница 186: ...9 5 6 4 MSDS SDS UV 1 Sensor setup 2 3 Maintenance 4 Hold 5 6 7 a 10 b 186...
Страница 188: ...5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 Diagnostics Test Signals Diagnostics Test Signals 188...
Страница 193: ......