Option
Description
Factor
Enters a slope factor to adjust the standard calibration.
Options: from 0.8 to 1.2 (default: 1)
Note: Use grab samples with known values or standard solutions to determine the slope
factor.
Standard
calibration
Start calibration
—Starts a one-point calibration.
Standard value
—Selects the value for the calibration standard.
Note: The Standard calibration option is not available if matrix calibration is active.
Calibration
interval
Sets the calibration interval. Options: off (default), 1 week, 4 weeks, 3 months or 6 months.
Note: Not available if matrix calibration is active.
5.3.2 Matrix calibration
The matrix calibration is only available for the NT3200sc probe and can have one, two or three
calibration points. The matrix calibration adapts the calculation for nitrate and nitrite to the conditions
in the medium. A matrix calibration is based on a laboratory comparison measurement. Start the
matrix calibration after the probe is in the medium and has adapted to the environment. Wait a
minimum of 15 minutes with the default configuration settings to let the probe adjust to the
temperature, complete some wiping cycles to clean the measurement gap and remove air bubbles.
The matrix calibration is necessary to adjust the measurements to the background environment. If a
single calibration point is used, it is recommended to select the calibration point at the lower limit of
the operating range of the application. For two calibration points, select the calibration points at the
lower and upper limit of the operating range. If a third calibration point is used, set the third
calibration point in the middle of the operating range.
The limit of the operating range depends on the application:
• If the nitrate concentration changes and nitrite is almost constant, set the limit to the lowest and
highest NO
3
concentration.
• If the nitrite concentration changes and nitrate is almost constant, set the limit to the lowest and
highest NO
2
concentration.
• If the nitrate and nitrite concentrations change, set the limit to the lowest and highest NO
3
concentration.
If one more calibration point is added, the probe knows automatically which of the three calibration
points needs to be rewritten.
Each calibration point is shown in
Matrix calibration > Information
and can be deleted manually.
If one of the events that follows occurs, delete all of the existing calibration points and record new
calibration points:
• The probe is moved to a different environment
• The matrix changes significantly
• The matrix changes seasonal
• The process changes
• Comparison measurements with laboratory values are not matching
• An error message shows that a matrix calibration is necessary
Note: Doing or erasing a matrix calibration does not have an effect on the factory calibration.
Make sure that the steps are followed for each calibration procedure:
• The probe is in the medium and has adapted to the environment.
• Collect the grab sample, then immediately validate on the controller that the grab sample was
collected.
• Collect the sample as near as possible to the sensor.
• Start the matrix calibration only when the concentrations are stable.
• To measure the grab sample, refer to
Instructions for laboratory measurements
16
English
Содержание NT3100sc
Страница 2: ...Table of Contents English 3 Fran ais 27 Espa ol 52 Portugu s 78 102 124 147 169 2...
Страница 11: ...Figure 3 Bracket installation English 11...
Страница 12: ...Figure 4 Flow direction top view Figure 5 Sensor with flow through unit installation 12 English...
Страница 20: ...Figure 8 Replace wiper blade 1 mm and 2 mm Figure 9 Replace wiper blade 5 mm 20 English...
Страница 36: ...Figure 4 Sens du d bit vue du dessus Figure 5 Installation d un capteur avec unit de d bit 36 Fran ais...
Страница 61: ...Figura 4 Direcci n del flujo vista superior Figura 5 Instalaci n del sensor con celda de flujo Espa ol 61...
Страница 86: ...Figura 3 Instala o do suporte 86 Portugu s...
Страница 87: ...Figura 4 Dire o do fluxo vista superior Figura 5 Instala o do sensor com a unidade de fluxo direto Portugu s 87...
Страница 104: ...2 1 2 1 1 2 1 2 104...
Страница 108: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5 mm 5 2 9 1 mm 2 mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 4 9 108...
Страница 109: ...3 109...
Страница 110: ...4 5 110...
Страница 118: ...9 5 mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 118...
Страница 126: ...2 1 1 2 1 2 126...
Страница 130: ...2 6 2 2 1 NT3100sc NT3200sc 3 5 mm 5 9 2 9 1 mm 2 mm 5 4 LZY261 3 3 1 9 130...
Страница 131: ...3 2 3 5 4 3 131...
Страница 132: ...4 5 132...
Страница 133: ...3 3 SC 6 6 4 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 133...
Страница 135: ...Off Auto Selection 3 6 12 24 1 3 1 2 3 4 5 3 5 3 1 1 SC200 SC1000 SC4500 Claros 2 3 Offset 10 10 0 124 0 8 1 2 1 135...
Страница 136: ...Start calibration Standard calibration 1 4 3 6 5 3 2 NT3200sc 1 2 3 15 1 2 3 3 NO3 NO2 NO3 1 3 146 1 1 2 3 3 1 2 ID 136...
Страница 140: ...8 1 mm 2 mm 9 5 mm 140...
Страница 142: ...10 7 5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 UNSTABLE 142...
Страница 143: ...5 6 NO3 NO3 N NO3 NO3 N NO3 NO3 N NO2 NO2 N NO2 NO2 N NO2 NO2 N 8 Web 143...
Страница 149: ...2 1 1 2 1 2 UV 149...
Страница 151: ...2 3 NT3100sc NT3200sc 2 4 250nm 2 5 NT3100sc NT3200sc 1 TSS3 SC SC 1 2 1 NT3100sc 1mm 2mm 5mm 5 000mg TSS L 3 TSS 151...
Страница 153: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5mm 5 2 9 1mm 2mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 9 153...
Страница 154: ...4 3 154...
Страница 155: ...4 5 155...
Страница 156: ...3 3 SC 6 6 4 ON 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 156...
Страница 158: ...3 6 12 24 1 3 1 2 3 4 5 3 5 3 1 1 SC200 SC1000 SC4500 Claros 2 3 10 10 0 147 0 8 1 2 1 1 1 4 3 6 158...
Страница 163: ...9 5mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 163...
Страница 164: ...c 10mL d 8 SC200 SC1000 SC4500 Claros Enter 9 NT3200sc NT3200sc NT3200sc 10 11 12 10 7 5 164...
Страница 165: ...5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 6 165...
Страница 171: ...5 1 2 PUR HT CE CMIM UKCA FCC ISED 1 2 2 2 1 2 1 1 171...
Страница 172: ...2 1 2 UV 2 1 3 2 1 4 ICES 003 Class A Class A 172...
Страница 175: ...1 1 3 2 2 6 2 2 1 NT3100sc NT3200sc 3 9 5 5 2 9 1 2 5 4 LZY261 9 175...
Страница 176: ...3 3 1 OFF 3 2 3 5 4 176...
Страница 177: ...3 177...
Страница 178: ...4 5 178...
Страница 186: ...9 5 6 4 MSDS SDS UV 1 Sensor setup 2 3 Maintenance 4 Hold 5 6 7 a 10 b 186...
Страница 188: ...5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 Diagnostics Test Signals Diagnostics Test Signals 188...
Страница 193: ......