Table 5 Error messages (continued)
Message
Possible cause
Solution
Wiper failed
The wiper is blocked.
The wiper position is not detected.
The wiper is defective.
Examine the measuring path and clean if
necessary.
Do a wiper test.
Contact technical support.
Flash lamp failed
The flash lamp or flash lamp control
is defective.
Contact technical support.
Humidity
There is too much humidity in the
sensor.
Remove the probe from the basin or flow-through
unit. Examine the humidity value in
Diagnostics/Test > Signals.
Contact technical support.
Temperature is out of
range
The temperature in the sensor is too
high.
Check the temperature in Diagnostics/Test >
Signals.
Check the environmental conditions.
Cool the sensor.
Try a different installation location.
Contact technical support.
Table 6 Warning and reminders
Message
Possible cause
Solution
Humidity
The humidity in the sensor is high.
Remove the probe from the basin or flow-
through unit. Examine the humidity value in
Diagnostics/Test > Signals.
Temperature is out of
range
The temperature in the sensor is too
high.
Check the temperature in Diagnostics/Test
> Signals.
Check the environmental conditions.
Cool the sensor.
Try a different installation location.
Contact technical support.
Wiper replacement
The time interval for servicing the wiper
blade has expired.
Replace the wiper blade.
Shaft seals
The time interval for servicing the shaft
seals has expired.
Contact technical support.
Seals
The time interval for servicing the seals
has expired.
Contact technical support.
Service
The time interval for factory service has
expired.
Contact technical support.
Calibration NO3/NO3N
The set calibration interval has expired.
Complete a standard calibration.
Calibration NO2/NO2N
The set calibration interval has expired.
Complete a standard calibration.
Matrix calibration
The set calibration interval has expired.
Complete a matrix calibration.
Section 8 Replacement parts and accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
English
23
Содержание NT3100sc
Страница 2: ...Table of Contents English 3 Fran ais 27 Espa ol 52 Portugu s 78 102 124 147 169 2...
Страница 11: ...Figure 3 Bracket installation English 11...
Страница 12: ...Figure 4 Flow direction top view Figure 5 Sensor with flow through unit installation 12 English...
Страница 20: ...Figure 8 Replace wiper blade 1 mm and 2 mm Figure 9 Replace wiper blade 5 mm 20 English...
Страница 36: ...Figure 4 Sens du d bit vue du dessus Figure 5 Installation d un capteur avec unit de d bit 36 Fran ais...
Страница 61: ...Figura 4 Direcci n del flujo vista superior Figura 5 Instalaci n del sensor con celda de flujo Espa ol 61...
Страница 86: ...Figura 3 Instala o do suporte 86 Portugu s...
Страница 87: ...Figura 4 Dire o do fluxo vista superior Figura 5 Instala o do sensor com a unidade de fluxo direto Portugu s 87...
Страница 104: ...2 1 2 1 1 2 1 2 104...
Страница 108: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5 mm 5 2 9 1 mm 2 mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 4 9 108...
Страница 109: ...3 109...
Страница 110: ...4 5 110...
Страница 118: ...9 5 mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 118...
Страница 126: ...2 1 1 2 1 2 126...
Страница 130: ...2 6 2 2 1 NT3100sc NT3200sc 3 5 mm 5 9 2 9 1 mm 2 mm 5 4 LZY261 3 3 1 9 130...
Страница 131: ...3 2 3 5 4 3 131...
Страница 132: ...4 5 132...
Страница 133: ...3 3 SC 6 6 4 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 133...
Страница 135: ...Off Auto Selection 3 6 12 24 1 3 1 2 3 4 5 3 5 3 1 1 SC200 SC1000 SC4500 Claros 2 3 Offset 10 10 0 124 0 8 1 2 1 135...
Страница 136: ...Start calibration Standard calibration 1 4 3 6 5 3 2 NT3200sc 1 2 3 15 1 2 3 3 NO3 NO2 NO3 1 3 146 1 1 2 3 3 1 2 ID 136...
Страница 140: ...8 1 mm 2 mm 9 5 mm 140...
Страница 142: ...10 7 5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 UNSTABLE 142...
Страница 143: ...5 6 NO3 NO3 N NO3 NO3 N NO3 NO3 N NO2 NO2 N NO2 NO2 N NO2 NO2 N 8 Web 143...
Страница 149: ...2 1 1 2 1 2 UV 149...
Страница 151: ...2 3 NT3100sc NT3200sc 2 4 250nm 2 5 NT3100sc NT3200sc 1 TSS3 SC SC 1 2 1 NT3100sc 1mm 2mm 5mm 5 000mg TSS L 3 TSS 151...
Страница 153: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5mm 5 2 9 1mm 2mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 9 153...
Страница 154: ...4 3 154...
Страница 155: ...4 5 155...
Страница 156: ...3 3 SC 6 6 4 ON 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 156...
Страница 158: ...3 6 12 24 1 3 1 2 3 4 5 3 5 3 1 1 SC200 SC1000 SC4500 Claros 2 3 10 10 0 147 0 8 1 2 1 1 1 4 3 6 158...
Страница 163: ...9 5mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 163...
Страница 164: ...c 10mL d 8 SC200 SC1000 SC4500 Claros Enter 9 NT3200sc NT3200sc NT3200sc 10 11 12 10 7 5 164...
Страница 165: ...5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 6 165...
Страница 171: ...5 1 2 PUR HT CE CMIM UKCA FCC ISED 1 2 2 2 1 2 1 1 171...
Страница 172: ...2 1 2 UV 2 1 3 2 1 4 ICES 003 Class A Class A 172...
Страница 175: ...1 1 3 2 2 6 2 2 1 NT3100sc NT3200sc 3 9 5 5 2 9 1 2 5 4 LZY261 9 175...
Страница 176: ...3 3 1 OFF 3 2 3 5 4 176...
Страница 177: ...3 177...
Страница 178: ...4 5 178...
Страница 186: ...9 5 6 4 MSDS SDS UV 1 Sensor setup 2 3 Maintenance 4 Hold 5 6 7 a 10 b 186...
Страница 188: ...5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 Diagnostics Test Signals Diagnostics Test Signals 188...
Страница 193: ......