Page 78
Bottle Mode – PAT700 Total Organic Carbon Analyzer
Operator Manual
Anatel
WG
M - 2
8
No
vemb
e
r 20
07
- Ed
it
io
n 0
1
The recommended correlation coefficient, a user-accepted value (R
2
), is 0.990 or greater.
The calibration will be accepted if the R
2
value is 0.990 or greater and the percent slope
change is within the specified
±
15% limit.
TOC calibration setup
Figure 6-3 TOC calibration setup dialog box
Use default settings and use custom settings radio buttons
enable you to select default
settings or custom settings. The default selection is use default settings.
Back icon
takes you back to the previous dialog box. All changes made to the TOC
calibration are retained.
Forward icon
takes you to the next dialog box.
•
If you select default settings, the load bottles dialog box opens.
.
•
If you select custom settings, the custom setup dialog box opens. See
. If
you select custom settings and then touch the back button, the analyzer uses the
default settings.
•
All changes made to the TOC calibration are retained.
Cancel icon
returns you to the run standards dialog box. All changes made in any of the
dialog boxes are lost.
Содержание ANATEL PAT700
Страница 1: ...ANATEL PAT700 ON LINE TOTAL ORGANIC CARBON ANALYZER Operator Manual...
Страница 2: ......
Страница 4: ......
Страница 56: ...Page 46 Run Modes PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 112: ...Page 102 Bottle Mode PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 136: ...Page 126 Data Review PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 154: ...Page 144 Specifications PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 156: ......
Страница 157: ......