WG
M - 2
8
No
vemb
e
r 20
07
- Ed
it
io
n 0
1
PAT700 Total Organic Carbon Analyzer – Diagnostics
Page 115
Anatel
Operator Manual
Digital I/O test
The digital I/O test icon allows you to test digital I/O communication. See
you perform the digital I/O test, make sure I/O wiring is properly connected.
1)
Connect the digital inputs to a system that is capable of sending a digital signal
2)
Configure one digital input with a high signal and the other with a low signal.
3)
Connect the digital outputs to a system that is capable of reading digital signals.
4)
Press the start test icon.
Figure 8-6 Digital I/O test dialog box
Start test icon
sends send a digital signal out the specified digital output port and reads the
signal on one of the digital input ports. If the read data matches the sent data, the test passes.
Return icon
returns you to the diagnostics dialog box.
4-20 mA output test
The 4-20 mA output test icon allows you to test the 4-20 mA outputs. See
. To
perform the test, you must connect a digital multimeter to the appropriate output to measure
the voltage.
1)
Connect a digital multimeter to one of the 4-20 mA outputs.
2)
Press the corresponding output icon to start the test.
Содержание ANATEL PAT700
Страница 1: ...ANATEL PAT700 ON LINE TOTAL ORGANIC CARBON ANALYZER Operator Manual...
Страница 2: ......
Страница 4: ......
Страница 56: ...Page 46 Run Modes PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 112: ...Page 102 Bottle Mode PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 136: ...Page 126 Data Review PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 154: ...Page 144 Specifications PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 156: ......
Страница 157: ......